What to do if...
Multi 3430
90
ba75865e02
12/2009
6.2
pH
Error message
OFL, UFL
Error message,
Error
Cause
Remedy
IDS pH sensor:
– Measured value outside the
measuring range
– Use suitable IDS pH sensor
– Air bubble in front of the dia-
phragm
– Remove air bubble
– Air in the diaphragm
– Extract air or moisten dia-
phragm
– Cable broken
– Exchange IDS pH sensor
– Gel electrolyte dried out
– Exchange IDS pH sensor
Cause
Remedy
IDS pH sensor:
– The values determined for zero
point and slope of the IDS pH
sensor are outside the allowed
limits.
– Recalibrate
– Junction contaminated
– Clean the junction
– IDS pH sensor broken
– Exchange IDS pH sensor
– Buffer solutions:
– Wrong buffer solutions
– Change calibration proce-
dure
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Change solutions