MIQ/Blue PS
Installation
3 - 5
ba75573e04
05/2009
Additional terminating
resistor on the MIQ/
Blue PS
An additional terminating resistor is only required if, except in the MIQ/
Blue PS, no other terminator switch is available on the IQ S
ENSOR
N
ET
island. The terminating resistor is connected to one of the three
SENSORNET connectors (Fig. 3-4). The terminating resistor must
bridge the two outside terminals of the SENSORNET connector. The
SENSORNET connector can still be used to connect a sensor.
Fig. 3-4
MIQ/Blue PS with additional terminating resistor
Examples of the correct
setting of the terminator
switches
The figure on the following page demonstrates with several examples
of IQ S
ENSOR
N
ET
islands how the terminator switches and terminating
resistors are correctly set.
6
8
7
9
2
1
3
5
4
SN terminator
switch ON
additional terminating resistor
X13
X12
100...
240V AC
NETZ/MAINS
L1
N
X11
X10
24V DC
EINGANG
INPUT
POWER
SENSORNET 2
SENSORNET 1
ON
EIN
OFF
AUS
X6
RO
T
RED
RO
T
RED
SN
TERMINA
TO
R
SCHIRM
SHIELD
GRÜN
GREEN
SCHIRM
SHIELD
GRÜN
GREEN
X5
X4
X3
X2
X1
SENSORNET 3
X8
X9
X7
RO
T
RED
SCHIRM
SHIELD
GRÜN
GREEN
+
-
R
R
R
connection selectable
Summary of Contents for MIQ/Blue PS SET
Page 30: ...Installation MIQ Blue PS 3 20 ba75573e04 05 2009...
Page 32: ...Maintenance and cleaning MIQ Blue PS 4 2 ba75573e04 05 2009...
Page 38: ...Technical data MIQ Blue PS 6 4 ba75573e04 05 2009...
Page 44: ...Appendix Installation instructions for several radio links MIQ Blue PS 7 6 ba75573e04 05 2009...
Page 45: ......