Installation
MIQ/24V
3 - 2
ba64114e02
08/2006
Note
Battery systems should be protected against total discharge. The
MIQ/24V does not have an integrated protection against total dis-
charge.
General installation
instructions
Follow these instructions when connecting the connecting wires to the
terminal strip:
!
Shorten all the wires to be used to the length required for the instal-
lation
!
Always fit all the ends of the wires with wire end sleeves before con-
necting them to the terminal strip
!
Any wires that are not used and project into the housing must be cut
off as closely as possible to the cable gland.
Warning
No free wires must be allowed to project into the housing. Other-
wise, there is a danger of electric short circuit that may cause fire.
Always cut off any wires that are not in use as closely as possible
to the cable gland.
Materials required
!
Wire end sleeves, suitable for the power line, with suitable crimping
tool
!
1 x cable gland with sealing ring (included in scope of delivery of
MIQ/24V).
Tools
!
Cable stripping knife
!
Wire stripper
!
Phillips screw driver
!
Small screw driver.
Preparing the
24 V AC/DC line
Fig. 3-1
Prepared 24 V AC/DC cable.
1
Cut off the cable to the required length.
2
Strip the cable insulation for approx. 45 mm.
3
Bare the wires 1 and 2 and fit them with wire end sleeves.
ca. 45 mm
wire 1
wire 2