MIQ/2-PR
Interface for the PROFIBUS
11
ba75974e01
07/2012
3
Interface for the PROFIBUS
The MIQ/2-PR PROFIBUS module enables to connect the
IQ S
ENSOR
N
ET
with all the sensors connected to it to the digital con
-
nection of a superordinate process control.
3.1
Scope of delivery
The scope of delivery of the MIQ/2-PR module comprises:
MIQ/2-PR
Accessory set, including:
– 4 x cable glands (clamping range 4.5-10 mm) with seals and blind
plugs
– 4x ISO blind nuts M4 with suitable cheese-head screws and plain
washers
– 2x countersunk screws M3x8 to close the module lid
(+ 2 replacement screws)
– 1 x contact base with fixing screws
Operating manual.
3.2
PROFIBUS checklist
The following checklist supports you when planning, projecting and
installing a PROFIBUS plant with the IQ S
ENSOR
N
ET
. For smooth
operation, you should be able to answer all questions with "Yes".
Is the [bus system] (bus segment) installed without branch lines?
Was the correct PROFIBUS cable used (cord type A according to
EN 50170)?
Are the poles of the signal lines A line and B line correctly connected
at all bus connections?
Is it guaranteed that there is no short-circuit between A line, B line
and cable shielding?
Is the shielding installed free of interruptions?
Are the guidelines for shielding and grounding being observed, and
doesn't any unallowed potential equalization current flow
via the shielding?
Further instructions for the installation and operation of a
PROFIBUS network is given on the Internet under
www.profibus.com and www.hms-networks.com.
Summary of Contents for MIQ/2-PR
Page 4: ...Contents MIQ 2 PR 4 ba75974e01 07 2012...
Page 10: ...Safety MIQ 2 PR 2 10 ba75974e01 07 2012...
Page 22: ...Interface for the PROFIBUS MIQ 2 PR 22 ba75974e01 07 2012...
Page 24: ...What to do if MIQ 2 PR 24 ba75974e01 07 2012...
Page 26: ...Technical data MIQ 2 PR 26 ba75974e01 07 2012...
Page 27: ......