System 2020 XT C
ONNECT
Installation
3 - 13
ba75970e01
10/2012
3.5
Connecting system components
3.5.1 General information
The IQ S
ENSOR
N
ET
system components are connected to form a func
-
tioning unit in the following ways:
Stacked mounting of MIQ modules
Up to three MIQ modules can be installed and mechanically con
-
nected with one another to form a stack at a single location. The con
-
tacts on the front and rear automatically create the electrical
connection between the MIQ modules in the stack (section 3.5.2).
Distributed mounting of MIQ modules
Individual MIQ modules or an assembled module stack are installed
at various locations (for example, a central control unit and an
MIQ branching module for connecting IQ sensors at the edge of the
basin). The spatial distance is bridged with the aid of the SNCIQ or
SNCIQ/UG cable (section 3.5.3).
Connection of IQ sensors:
IQ sensors are connected to an MIQ module with the aid of the
SACIQ cable. This cable is available in different lengths. The
MIQ module must have a free SENSORNET connection on the ter
-
minal strip. On the sensor side, the connection is made via a plug
that is standard for all IQ sensors (section 3.5.4).
Docking of a terminal unit:
A terminal unit is attached to the lid of a free MIQ module with the
aid of a quick-release locking mechanism.
Note
Distributed mounting is only allowed in a star configuration. The system
must nowhere be closed to form a ring. Only use the assembly material
provided for the installation.
Summary of Contents for IQ SENSOR NET System 2020 XT
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 8: ...Contents System 2020 XT CONNECT 0 6 ba75970e01 10 2012...
Page 24: ...Overview System 2020 XT CONNECT 1 16 ba75970e01 10 2012...
Page 28: ...Safety System 2020 XT CONNECT 2 4 ba75970e01 10 2012...
Page 123: ...System 2020 XT CONNECT Settings setup 5 11 ba75970e01 10 2012 Fig 5 10 Edit list of sensors...
Page 138: ...Settings setup System 2020 XT CONNECT 5 26 ba75970e01 10 2012...
Page 152: ...Network connection System 2020 XT CONNECT 6 14 ba75970e01 10 2012...
Page 178: ...PROFIBUS connection option System 2020 XT CONNECT 8 12 ba75970e01 10 2012...
Page 180: ...Maintenance and cleaning System 2020 XT CONNECT 9 2 ba75970e01 10 2012...
Page 194: ...What to do if System 2020 XT CONNECT 10 14 ba75970e01 10 2012...
Page 204: ...Accessories and options System 2020 XT CONNECT 12 2 ba75970e01 10 2012...
Page 212: ...Lists IQ SENSOR NET System 2020 XT CONNECT 13 8 ba75970e01 10 2012...
Page 215: ......