3.5.2 Stacked mounting of MIQ modules:
Note
For optimum stability, a maximum of three MIQ modules can be mount
-
ed in any one stack. Only one MIQ power supply module may be
mounted per module stack.
Mounting direction
MIQ modules can be stacked on top of one another from both sides. All
the required installation materials are contained in the scope of delivery
of the MIQ modules.
Which variant is suitable for a specific case depends on whether and
how an MIQ module is already permanently installed at a location (e.g.
on a wall or in a panel), among other factors.
Materials required
2 x ISO blind nuts M4
2 x cheese-head screws M4x16 with plastic washer
1 x contact base with two plastic tapping screws.
Tools
Phillips screw driver.
Variant 1 - stack expansion for
-
wards.
The back of an MIQ module is
attached to the lid of another
MIQ module
Variant 2: - stack expansion
backwards.
The lid of an MIQ module is
attached to the back of another
MIQ module (page 3-18).
In the case of panel mounting, the front MIQ module can
only be installed individually in the switch cabinet aperture.
Only then can further MIQ modules be added to its back
(variant 2). If necessary, dismantle the front MIQ module
from an already mounted stack before mounting.
Summary of Contents for IQ SENSOR NET System 2020 XT
Page 2: ...Copyright 2016 Xylem Analytics Germany GmbH Printed in Germany...
Page 8: ...Contents System 2020 XT CONNECT 0 6 ba75970e01 10 2012...
Page 24: ...Overview System 2020 XT CONNECT 1 16 ba75970e01 10 2012...
Page 28: ...Safety System 2020 XT CONNECT 2 4 ba75970e01 10 2012...
Page 123: ...System 2020 XT CONNECT Settings setup 5 11 ba75970e01 10 2012 Fig 5 10 Edit list of sensors...
Page 138: ...Settings setup System 2020 XT CONNECT 5 26 ba75970e01 10 2012...
Page 152: ...Network connection System 2020 XT CONNECT 6 14 ba75970e01 10 2012...
Page 178: ...PROFIBUS connection option System 2020 XT CONNECT 8 12 ba75970e01 10 2012...
Page 180: ...Maintenance and cleaning System 2020 XT CONNECT 9 2 ba75970e01 10 2012...
Page 194: ...What to do if System 2020 XT CONNECT 10 14 ba75970e01 10 2012...
Page 204: ...Accessories and options System 2020 XT CONNECT 12 2 ba75970e01 10 2012...
Page 212: ...Lists IQ SENSOR NET System 2020 XT CONNECT 13 8 ba75970e01 10 2012...
Page 215: ......