pH
73
6.1.3
Measured value resolution of voltage display
Using the U-AutoRng function, you can select the resolution
of the voltage measurement:
l
If U-AutoRng is not active, voltages in the range of -1999
to +1999 mV are measured with a resolution of 1 mV.
l
If U-AutoRng is active, voltages in the range of
+999.9 mV to -999.9 mV are measured with a resolution
of 0.1 mV.
The rest of the measuring range remains unchanged.
Note
If you expect voltages outside the range +999.9 mV to -
999.9 mV and want to connect a recorder, deactivate U-
AutoRange. Otherwise, the recorder scaling at the range
limits will change.
Measured value
resolution,
U
1
Switch to the
Menu
mode.
2
Use
4 6
to select the main
Meas
menu.
3
Select the
pH setup
menu item with
82
.
4
Confirm the
pH setup
menu item with
g
.
5
Select the
U-AutoRng
menu item with
82
.
6
Confirm the
U-AutoRng
menu item with
g
.
The
U-AutoRng
function is active (
✔
).
Voltages in the range between -999.9 and +999.9 are
displayed with a resolution of 0.1 mV.
7
Terminate the
Menu
mode with
m
.