Determining/Selecting the cell constant C
Cond 3150i
6
Determining/Selecting the cell constant C
The cell constant of the conductivity measuring cell can be set to one of the
following values:
z
0.475 cm
-1
z
0.100 cm
-1
z
0.875 cm
-1
Note
The cell constant that should be selected must either be taken from the op-
erating manual of the measuring cell or is printed on the measuring cell.
Procedure for setting up the cell constant:
z
Press the
<CAL>
key repeatedly until the
CELL
display appears.
z
Then press the
<RUN/ENTER>
key.
z
Press the
<CAL>
key repeatedly until the required cell constant appears
in the lower line of the display.
Selection: 0.475 cm
-1
; 0.100 cm
-1
; 0.875 cm
-1
z
Switch to the measuring mode with
<M>
.
L
C L
E
5
0 7
4
2 5
9
cm
1/
S/
cm
Tref25
nLF
ARng
Summary of Contents for Cond 3150i
Page 12: ...Technical data Cond 3150i 12...