pH 298, LF 298, Oxi 298, Cl 298
Outputs
49
ba77196e03
07/2018
5.3.2 Configuring the analog outputs
1
Open the
configuration
menu with
<CONF>
.
2
Using
<
>
, highlight the
PID controller
menu item and confirm
with
<ENTER>
.
3
Select the setting
release
/
off
or
type of the controller
/
PWM
.
4
Quit the menu
PID controller
.
1
Switch off the
PID controller
(
release
/
off
)
or configure it as a pulse width
controller (
type of the controller
/
PWM
)
(see section 5.2.2).
2
Open the
configuration
menu with
<CONF>
.
3
Using
<
>
, highlight the
outputs
menu item and confirm with
<ENTER>
.
Submenu
Selection/Values
Explanation
output 1
output 2
current output
output val.
measured value
2nd value
temperature
val. 0/4 mA
Input field for the 0/4 mA start value
val. 20 mA
Input field for the 20 mA end value
type
0...20 mA
4...20 mA
type u/i
current
voltage
Select the analog output form
current
= 2 x 0/4…20 mA
voltage
= 2 x 0…5 V
This configuration has to agree with
the hardware coding (delivery set-
ting:
current
= 2 x 0/4…20 mA).
The output as voltage has to be
coded by setting a jumper on the
hardware.
4
Configure the outputs with
<
>
and
<ENTER>
.
5
Quit the menu.