Operation
pH 298, LF 298, Oxi 298, Cl 298
32
ba77196e03
07/2018
For this the temperature of the measurement and calibration media has to be
measured or stored as a fixed temperature. If the fixed temperature is enabled
it is used for temperature compensation. The Fix message appears in the upper
right-hand corner of the display.
Setting the fixed
temperature
4.7
Calibrating (CAL menu)
4.7.1 General proceeding
General
proceeding
1
Open the
configuration
menu with
<CONF>
.
2
Using
<
>
, highlight the
temperature
menu item and confirm
with
<ENTER>
.
Submenu
Selection/Val-
ues
Explanation
fixed temp.
on
off
Enabling / disabling the fixed tempera-
ture
temp. Value
Input field for the fixed temperature value
3
Select the setting
fixed temp.
/
on
.
4
Set the
temp. Value
in the input field.
5
Quit the menu
configuration
.
The sensor is protected by a password against unauthorized cali-
bration. For more information on password protection, refer to
section 4.11.
1
Switch to the measured value display as necessary.
2
Open the
calibration
menu with
<CAL>
.
The
password
input field pops up.
3
Enter the password.
The
calibration
menu is open.
The
calibration
menu lists the calibration procedures suitable for the
instrument type.
Details on calibration for your type of instrument are given in the fol-
lowing sections (see section 4.7.3 ... section 4.7.6).