AmmoLyt
®
Plus
700 IQ
Commissioning
3 - 13
ba75651e03
08/2009
3.5.3
Setting table of the
K
(potassium sensor)
Note
These sensor settings are only available if in the sensor settings for the
the
K
setting was set to
active
The basic settings are taken over from the sensor but can
be adjusted separately afterwards.
Menu item
Selection/values
Explanations
Measuring mode
z
K (mg/l)
z
mV
Citation form of the mass concentration or
voltage of the electrode.
Measuring range
(
Measuring mode
:
K (mg/l)
)
z
1 ... 1000 mg/l
Fixed range
Measuring range
(
Measuring mode
:
mV
)
z
-2000 ... 2000 mV
Fixed range
IQ-LabLink
z
active
z
not active
active
(standard setting):
During the matrix adjustment, operating
steps for the data transfer with the IQ-
LabLink procedure by means of a USB
memory device are offered (see section
4.2.3)
not active
:
During the matrix adjustment, operating
steps for manual entry are offered.
Temperature mode
(only with
Measuring mode
:
K (mg/l)
)
z
°C
z
°F
Unit of the measured temperature value
(Celsius, Fahrenheit).