Commissioning
AmmoLyt
®
Plus
700 IQ
3 - 10
ba75651e03
08/2009
3.5.2
Setting table of the
Menu item
Selection/values
Explanations
Measuring mode
z
NH4-N
z
NH4
z
mV
Citation form of the mass concentration or
voltage of the electrode.
Measuring range
(
Measuring mode
:
NH4-N
)
z
AutoRange
z
0.1 ... 100.0 mg/l
z
1 ... 1000 mg/l
2 measuring ranges can be selected.
With
AutoRange
, the instrument
automatically switches to the suitable
measuring range.
Measuring range
(
Measuring mode
:
NH4
)
z
AutoRange
z
0.1 ... 129.0 mg/l
z
1 ... 1290 mg/l
2 measuring ranges can be selected.
With
AutoRange
, the instrument
automatically switches to the suitable
measuring range.
Measuring range
(
Measuring mode
:
mV
)
z
-2000 ... 2000 mV
Fixed range
K
z
hide
z
display
hide
(standard setting):
The potassium electrode is not displayed
as an extra sensor in the measured value
display.
display
:
If the sensor is equipped with the
potassium electrode, the electrode is
displayed as the extra sensor
K
in the measured value display.
The relevant settings can be done in an
extra setting table (see section 3.5.3).
Note: If the maximum number of sensors
for the system is exceeded by activating
the sensor, the
K
cannot be
activated.
Calib. history K
(only with the setting
K
:
hide
)
z
Do not download
z
Transmit to log book
Transmit to log book
generates a log book
message with the calibration history of the
potassium electrode. When opening the
setting table again the setting is reset to
Do not download
.
Summary of Contents for AmmoLyt Plus 700 IQ
Page 12: ...Safety instructions AmmoLyt Plus 700 IQ 2 4 ba75651e03 08 2009...
Page 40: ...Matrix adjustment check and calibration AmmoLyt Plus 700 IQ 4 14 ba75651e03 08 2009...
Page 42: ...Measuring AmmoLyt Plus 700 IQ 5 2 ba75651e03 08 2009...
Page 48: ...Maintenance and electrode exchange AmmoLyt Plus 700 IQ 6 6 ba75651e03 08 2009...
Page 62: ...Indexes AmmoLyt Plus 700 IQ 10 4 ba75651e03 08 2009...