Operation of the Single Measuring System
OxiTop
Measuring principle
BOD measurement with the OxiTop
measuring system is based on pressure
measurement (difference measurement). The measuring is made by pressure
measurement via
piezoresistive electronic pressure sensors
.
With the following functions, the OxiTop
measuring system minimizes the measuring
work and is especially suited to the courses of the respirometric BOD measurement:
•
•
•
•
AUTO TEMP function:
Control of the
temperature adjustment
and
automatic start
of the measurement
(1 hour after starting the OxiTop
measuring system at the earliest, after 3 hours at
the latest). It is not necessary to temper the samples to exactly 20°C before starting
OxiTop
. Samples between 15°C and 20°C can be started immediately as BOD
measurement will not begin until a stable 20°C is reached. This is controlled by
the "AutoTemp" feature.
•
•
•
•
Data logging
Automatic measurement with daily storage of measured value for 5 days:
facilitates measurement without supervising, e. g. over the weekend.
•
•
•
•
Current value
Indication of measured value (0 - 40 digits) corresponds to the digits of the
previous WTW BOD instruments. Conversion of mbar to digits is stored in the
instrument, i. e. evaluation of the measured values remains unchanged.
•
•
•
•
Measuring range reserve
from 40 to 50 digits - for overranging without reset by opening the bottle.
Summary of Contents for 208210
Page 1: ......
Page 2: ......
Page 14: ...Bedienungsanleitung 14 Inductive Stirring System IS 6 IS 12 IS 12 6...
Page 19: ...Technische Daten 19 Pr fzertifikat EMV IS 6...
Page 20: ...Technische Daten 20 Pr fzertifikat EMV IS 12...
Page 21: ...Technische Kundenberatung...
Page 22: ......
Page 34: ...Instruction manual 34 Inductive Stirring System IS 6 IS 12 IS 12 6...
Page 39: ...Technical data 39 Test certificate EMC IS 6...
Page 40: ...Technical data 40 Test certificate EMC IS 12...
Page 41: ......
Page 42: ......
Page 43: ......