15
5.
Rinse with cold tap water for a minimum of one minute; use a syringe to repeatedly flush
any very narrow lumens.
6.
Bathe in a detergent solution prepared per manufacturer directions for 5 minutes.
7.
Scrub thoroughly with a soft brush and/or pipe cleaner; repeatedly flush any very narrow
lumens with detergent solution using a syringe.
8.
Rinse thoroughly /flush with deionized / reverse osmosis (RO/DI) water.
9.
Sonicate for a minimum of 10 minutes in an enzymatic detergent solution prepared per
manufacturer directions.
10.
Rinse thoroughly /flush with RO/DI water.
11.
Dry with a clean, soft, absorbent, disposable cloth.
12.
Visually inspect for cleanliness. All visible surfaces, internal and external, should be
visually inspected. If necessary re-clean until it is visibly clean.
Steam Sterilization
Cycle Type
Parameter
Minimum Set Point
Prevacuum
Exposure Temperature
270°F (132°C)
270°F (132°C)
Exposure Time
4 minutes
Dry Time
20 minutes