Insert a Locator Pin into the first hole to maintain proper reduction while the 
second hole is drilled.  Drill second hole to the proper depth using the laser 
marked Reamer. Remove the Reamer, Locator Pin and Reamer Guide. Remove 
the Inserter / Implant assembly from the sterile Kit. 

| Figure 3

 While maintaining reduction, align and insert the tips of the FUSEFORCE® into 
the drilled holes. 

| Figure 4

  

Tamp the FUSEFORCE® as flush to the bone as possible. Implant positioning 
may be evaluated radiographically.

 | Figure 5

| Figure 3

| Figure 4

| Figure 5

Chapter 2

     Surgical Technique

Summary of Contents for FUSEFORCE

Page 1: ... FUSEFORCE Hand Fixation System SURGICAL TECHNIQUE ...

Page 2: ... evaluate the appropriateness of the procedures based on his or her personal medical training and experience Prior to use of the system the surgeon should refer to the product package insert for complete warnings precautions indications contraindications and adverse effects Package inserts are also available by contacting Wright Medical Technology Inc Contact information can be found on the back o...

Page 3: ...increased potential for device failure when excessive demands are made upon it Strenuous loading excessive mobility and articular instability all may lead to accelerated wear and eventual failure by loosening fracture or dislocation of the device If excessive loading cannot be prevented an implant should not be used Abnormal or excessive forces could lead to delayed union non union or failure of t...

Page 4: ...tape measure Open the matching size FUSEFORCE Implant Kit with its corresponding Inserter Reamer Reamer Guide and Locator Pin Procedure While maintaining desired reduction place Reamer Guide across the fusion site with both guides touching bone Figure 1 Drill the first hole to the proper depth using the laser marked Reamer provided in the Kit Figure 2 Figure 1 Figure 2 ...

Page 5: ...ed Reamer Remove the Reamer Locator Pin and Reamer Guide Remove the Inserter Implant assembly from the sterile Kit Figure 3 While maintaining reduction align and insert the tips of the FUSEFORCE into the drilled holes Figure 4 Tamp the FUSEFORCE as flush to the bone as possible Implant positioning may be evaluated radiographically Figure 5 Figure 3 Figure 4 Figure 5 Chapter 2 Surgical Technique ...

Page 6: ...the plastic Inserter counter clockwise as shown by arrows on top to release the Implant Place the end of the Inserter on the implant and apply pressure manually or using a mallet to fully seat the FUSEFORCE Implant Chapter 2 Surgical Technique ...

Page 7: ...NS0808 FFS 0808 FUSEFORCE 8 8 FFNS1010 FFS 1010 FUSEFORCE 10 10 FFNS1012 FFS 1012 FUSEFORCE 10 12 FFNS1212 FFS 1212 FUSEFORCE 12 12 FFNS1515 FFS 1515 FUSEFORCE 15 15 FFNS1816 FFS 1816 FUSEFORCE 18 16 FFNS2020 FFS 2020 FUSEFORCE 20 20 FFNS2522 FFS 2522 FUSEFORCE 25 22 FFNS101513 FFS 101513 FUSEFORCE 10 15 13 B A 2mm 3mm 2mm B A 2mm 3mm 2mm B A 2mm 3mm 2mm B A 2mm 3mm 2mm ALT PN is the Solana part n...

Page 8: ...right Medical Technology Inc 1023 Cherry Road Memphis TN 38117 800 238 7117 901 867 9971 www wmt com Wright Medical EMEA Atlas Arena Australia Building Hoogoorddreef 7 1101 BA Amsterdam the Netherlands 011 31 20 565 9060 Wright Medical UK Ltd Unit 1 Campus Five Letchworth Garden City Hertfordshire SG6 2JF United Kingdom 011 44 0 845 833 4435 ...