Insert a Locator Pin into the first hole to maintain proper reduction while the
second hole is drilled. Drill second hole to the proper depth using the laser
marked Reamer. Remove the Reamer, Locator Pin and Reamer Guide. Remove
the Inserter / Implant assembly from the sterile Kit.
| Figure 3
While maintaining reduction, align and insert the tips of the FUSEFORCE® into
the drilled holes.
| Figure 4
Tamp the FUSEFORCE® as flush to the bone as possible. Implant positioning
may be evaluated radiographically.
| Figure 5
| Figure 3
| Figure 4
| Figure 5
Chapter 2
Surgical Technique