SI-MT/SI-MKB/SI-HTB
WORLD PRECISION IN STRU MENTS
31
Using the Sarcomere Length Unit
Setup
1. Turn on the system. See "Turning the System On" on page 25.
2. Line up the laser connector with the port labeled
Laser
on the
SI-SARCAM
module
and press it into place.
3. Line up the camera connector with the port labeled
CCD
on the
SI-SARCAM
module
and press it into place.
4. Press
the
Setup
button to toggle through the setup parameters.
5. Press
the
Display
button to save the confi guration and return to the normal display.
NOTE
: The unit remembers the state of all the parameters, even after it is powered
off. To reset the factory defaults, turn the unit off, press both the
Up
and
Down
but-
tons simultaneously while you turn the system back on.
NOTE
: For safety reasons, the laser is always disabled at startup.
Choosing a Display Mode
The default display is two lines and shows the calculated (exact) sarcomere spacing, fi rst
order diffraction distance that is used to calculate the sarcomere spacing and the signal
amplitude, which is used when you are aligning the camera. To toggle through the display
modes, press the
Display
button. Press one time to see the fi rst order diffraction distance.
Press it again to view the signal amplitude. Press it a third time to return to the sarcomere
spacing display.
Setup Menu
Press the Setup button to toggle through the Setup menu and cycle through the list of
available parameters. Parameters are shown in Fig. 26.
CCD Offset
CCD Offset
CCD Sensitivity (1–10)
CCD Sensitivity
Tissue to CCD Distance
Tissue Distance
CCD Mode (263Hz/1KHz)
CCD Mode
Backlight Level (1–8)
Backlight Level
Laser Off/On
LASER DISABLED
Fig. 26 The Setup button lets you toggle through the list of parameters.