SPECIFICATIONS

Single Channel Pipettes

Model

Volume

Range (μL)

Increment

(μL)

Nominal

Volume (μL)

Tolerance

%

Repeatability

%

 REG2

0.2 ~ 2

0.01

0.2
0.5

2

±12.0

±5.0
±2.0

6.00

2.50

0.70

 REG10

1 ~ 10

0.1

1
5

10

±3.0
±1.5
±1.0

1.50

0.60

0.40

 REG20

2 ~ 20

0.1

2

10
20

±3.0
±1.0
±1.0

1.50

0.50

0.30

 REG50

5 ~ 50

0.5

5

20
50

±2.0
±1.2
±1.0

1.50

0.40

0.20

 REG100

10 ~ 100

1

10
50

100

±2.0
±0.8
±0.8

0.50

0.30

0.15

 REG200

20 ~ 200

1

20

100
200

±2.0
±0.8
±0.8

0.50

0.30

0.15

 REG1000

100 ~ 1000

5

100
500

1000

±1.5
±0.8
±0.8

0.30

0.30

0.15

 REG5K

1000 ~ 5000

50

1000
2000
5000

±1.0
±0.7
±0.7

0.50

0.25

0.15

 REG10K

1000 ~ 10000

100

1 mL
5 mL

10 mL

±3.0
±0.7
±0.7

0.30

0.20

0.15

Summary of Contents for REG10

Page 1: ...INSTRUCTION MANUAL Serial No _____________________ 030813 Regal Pipettes World Precision Instruments www wpiinc com...

Page 2: ...ENTS ABOUT THIS MANUAL 1 INTRODUCTION 2 Cautions and Warnings 2 Parts List 2 Single Channel Pipettes 2 INSTRUMENT DESCRIPTION 3 Single Multi Channel Crystal Pipettes 3 Setting the Dispense Volume 4 OP...

Page 3: ...iv World Precision Instruments...

Page 4: ...ON Cautions warn against actions that can cause damage to equipment Please read these carefully This symbol indicates a WARNING Warnings alert you to actions that can cause personal injury or pose a p...

Page 5: ...the pipette down when a lled tip is attached CAUTION NEVER use a pipette without a tip attached Parts List After unpacking verify that there is no visible damage to the product Verify that all items...

Page 6: ...mm four inches of shock absorbing material For further details please read the section entitled Claims and Returns on page 11 of this manual INSTRUMENT DESCRIPTION Single Multi Channel Crystal Pipette...

Page 7: ...e nose cone Setting the Dispense Volume Rotate the adjustment dial or the push button to dial in the aliquot volume you want to dispense Rotate the dial to the left to decrease and to the right to inc...

Page 8: ...CAUTION Do not lay the pipette down when a lled tip is attached Dispensing Liquid 1 To dispense the liquid position the pipette tip close to the inner wall of the receiving vessel and press the push b...

Page 9: ...pipette to the maximum volume Place a disposable tip on the pipette Using a balance measure the weight of three successive aliquots of distilled water to the nearest 0 0001 of a gram 2 Record each val...

Page 10: ...pette Tips Universal Standard Universal Tips Tip Volume For Pipetter Bulk Part No Rack Part No 0 1 10 L REG2 REG10 REG20 Bag of 1000 500191 960 10 racks of 96 500192 5 200 L REG20 REG50 REG100 REG200...

Page 11: ...t No 0 1 10 L Flat 0 17 mm OD REG2 REG10 REG20 200 500202 0 1 10 L Flat 0 37 mm OD REG2 REG10 REG20 200 500203 0 1 10 L Round 0 57 mm OD REG2 REG10 REG20 200 500204 0 5 200 L Round 0 57 mm OD REG20 RE...

Page 12: ...1 5 10 3 0 1 5 1 0 1 50 0 60 0 40 REG20 2 20 0 1 2 10 20 3 0 1 0 1 0 1 50 0 50 0 30 REG50 5 50 0 5 5 20 50 2 0 1 2 1 0 1 50 0 40 0 20 REG100 10 100 1 10 50 100 2 0 0 8 0 8 0 50 0 30 0 15 REG200 20 20...

Page 13: ...10 World Precision Instruments...

Page 14: ...rect special or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly...

Page 15: ...44 0 1462 424700 Fax 44 0 1462 424701 E mail wpiuk wpi europe com Germany Zossener Str 55 10961 Berlin Tel 030 6188845 Fax 030 6188670 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading...

Reviews: