background image

MityFlex

World Precision Instruments 

3

OPERATING INSTRUCTIONS

Installation

POWER REQUIREMENTS: 

Voltage and frequency of power supply must be the same 

as shown on unit.

WIRING CONNECTIONS: 

All wiring and electrical connections must comply with local 

and national electrical codes.

MOUNTING: 

Place on any  flat surface. Ensure that space is provided at the back 

of the case for air circulation through ventilation holes. Do not allow tubing to be 

crimped.

MOUNTING LOCATION:

 Pump should be placed in a dry location with adequate 

cooling provided. The ambient  temperature should not exceed 40°C. 

This unit should 

not be used outdoors or in hazardous locations.

Roller/Tubing Selection

Use the following color coded rollers with the appropriate tubing I.D. Your unit is 

assembled with black rollers. Red rollers are included in the optional 

MITY-KIT

. See 

separate instructions for roller replacement enclosed in your tube kit.

 

Roller 

Tubing ID 

Tubing OD 

mL/Revolution

 Black 

1

/

4

” (0.250) 

7

/

16

” (0.438”/11.1mm) 

3.5

 Black 

3

/

16

” (0.187) 

3

/

8

” (0.375”/9.5mm) 

2.1

 Red 

1

/

8

” (0.125) 

*

 

3

/

8

” (0.375”/9.5mm) 

0.8

 Red 

1

/

16

” (0.062)

*

 

3

/

8

” (0.375”/9.5mm) 

0.21

* Red rollers for use with 

1

/

8

” and 

1

/

16

” tubing are included in the optional MITY-KIT.

NOTE

: The 

1

/

16

” ID tubing is actually a small piece of tubing inside a larger piece.

Variable Speed

Variable speed control may be set at any level before turning unit on.
Unit is designed for continuous operation at any speed.

CAUTION: 

Avoid prolonged operation in the stalled position to ensure longer 

motor life.

Summary of Contents for MityFlex

Page 1: ...Instrumenting scienti c ideas WORLD PRECISION INSTRUMENTS www wpiinc com INSTRUCTION MANUAL 051316 MityFlex Peristaltic Pump Serial No _____________________...

Page 2: ......

Page 3: ...orm without prior written permission of World Precision Instruments Inc CONTENTS ABOUT THIS MANUAL 1 INTRODUCTION 1 SPECIFICATIONS 2 OPERATING INSTRUCTIONS 3 Installation 3 Roller Tubing Selection 3 V...

Page 4: ...ii World Precision Instruments...

Page 5: ...that can cause personal injury or pose a physical threat Please read these carefully NOTES and TIPS contain helpful information INTRODUCTION MITYFLEX variable speed drive is powered by DC gearmotors...

Page 6: ...le speed drive system reversible FUSE Located in the inlet receptacle Uses 5 X 20mm 1 0 amp See page 4 for replacement procedure DC MOTOR BRUSHES Replaceable See page 4 for replacement procedure WEIGH...

Page 7: ...ons Roller Tubing Selection Use the following color coded rollers with the appropriate tubing I D Your unit is assembled with black rollers Red rollers are included in the optional MITY KIT See separa...

Page 8: ...pers see Fig 3 If using 9 piece of tube 11 2 should be extended outside of inlet port 8 Continue to rotate the roller bracket assembly using the extra bracket while pushing the tubing into the roller...

Page 9: ...ld fuse and replace with new fuse 6 Push drawer in until it locks in place Motor Brush Replacement 1 Disconnect power to pump 2 Remove case cover by removing 8 screws 3 With a flat head screw driver r...