WPI Bridge-8 Instruction Manual Download Page 16

12 

World Precision Instruments

Output Connections

The output from each Iso-DAM8A module is available on the rear panel from a 

standard BNC connector or all channels are available at the 25-pin connector. See 

Fig. 6 for pin designations. The Iso-DAM8A output is suitable for application to any 

recording or data acquisition device. 

Grounding

Proper grounding is critical to successful biological signal monitoring. The optimum 

grounding configuration is the “single-point ground,” in which each component of a 

monitoring setup is connected to a single point which is connected to earth or mains 

ground. This grounding configuration is used in 

Fig. 6 and Fig. 7.

Shielding

The isolation feature of the Iso-DAM8A minimizes, but cannot eliminate, noise pickup 

from electric distribution lines, fluorescent lights, computers, motors, radio frequency 

transmitters, microwave ovens, etc. Use of shielded wires for input connection is 

necessary if the subject is located more than a few inches from the amplifier inputs. 

Even the use of shielded wires and proper grounding may be insufficient in a high 

electrical noise environment, particularly if high impedance electrodes are used, if the 

signal level is low (less than 100mV or so), or if high frequency response is required. In 

these situations, the use of a metal shield, known as a Faraday cage, which encloses 

the subject and the headstage may be necessary. 

Idle Settings

If you are not certain what settings are required, the following settings are a good 

starting point:

PROBE SELECT: GND

LOW FILTER: DC

DC OFFSET: OFF

GAIN: 10

HIGH FILTER: 10 kHZ

NOTCH FILTER: OFF

POSITION: Mid-range

POWER: OFF

Control Adjustment

After connecting inputs and outputs, the amplifier can be turned on. The POSITION 

control may be used to establish a baseline setting on the recording device. It is 

normal to set this control so that the recording device indicates zero when the PROBE 

SELECT is in the GND position. The PROBE SELECT switch should be set appropriately 

for the setup. 

• 

If the headstage is used, set the switch to PROBE. 

• 

If direct connection is made to the Iso-DAM8A input, set the switch to INT.

Summary of Contents for Bridge-8

Page 1: ...WORLD PRECISION INSTRUMENTS ISDB ISO DAM 8A and BRIDGE8 Serial No _____________________ www wpiinc com INSTRUCTION MANUAL 060920...

Page 2: ......

Page 3: ...Output to Zero 3 Bridge Switch 3 Transducer Calibration 4 Example 4 General Amplification 4 Specifications 4 ISO DAM8A 6 Introduction 7 Notes and Warnings 7 Unpacking 7 Instrument Description 8 Noise...

Page 4: ...ii World Precision Instruments...

Page 5: ...on Fig 1 Up to eight rackmountable ISO DAM8A and Bridge8 modules can be combined in a single chassis Introduction Any combination of ISO DAM8A or Bridge 8 modules may be combined in the same instrumen...

Page 6: ...horized by collect freight may be refused If a return shipment is necessary use the original container if possible If the original container is not available use a suitable substitute that is rigid an...

Page 7: ...vel to nearly zero volts adjust the Offset Adjust knob above the Gain switch so that both lamps are lit together A significant increase in the sharpness or resolution of the zero adjustment may be ach...

Page 8: ...he output per mm Hg would be approximately 5 mV per mm Hg Since the nominal rating of 5 V per volt is usually not exact the user must in the end calibrate the transducer with a known hydrostatic press...

Page 9: ...arasota FL 34240 9258 USA As the manufacture distributor of the apparatus listed declare under sole responsibility that the product s WPI PN ISDB SERIES To which this declaration relates is are in con...

Page 10: ...at can cause damage to equipment Please read these carefully This symbol indicates a WARNING Warnings alert you to actions that can cause personal injury or pose a physical threat Please read these ca...

Page 11: ...set to 50 or 60 Hz and an output DC position adjustment The Iso DAM8 chassis can be ordered for either 110 120VAC or 220 240VAC operation The chassis can accommodate up to eight Iso DAM8A bioamplifier...

Page 12: ...he preparation and electrode lead wires in a Faraday cage or shield to reduce mains induction to sufficiently low levels It may be beneficial to extinguish overhead fluorescent lighting as well Use of...

Page 13: ...ust the Gain switch position stability This potentiometer is accessible through a small screwdriver hole on the front panel To adjust the DC balance connect the and inputs of the input connector or th...

Page 14: ...gly necessary if high gains are to be used with DC amplification The toggle switch below the DC Offset knob selects the polarity of the offset voltage If offset adjustment is not required the toggle s...

Page 15: ...rode i e TM33B01 EP2 Ag AgCl Half Cell Prep A B Central Ground 3294 Grounding Clip Fig 7 A typical single ended configuration is shown here The Iso DAM8A inputs can be connected in either differential...

Page 16: ...than a few inches from the amplifier inputs Even the use of shielded wires and proper grounding may be insufficient in a high electrical noise environment particularly if high impedance electrodes ar...

Page 17: ...erely distort some signals If a large DC offset is present in the setup the DC OFFSET control can be used to cancel it The setting of this control can be quite sensitive especially at high gain settin...

Page 18: ...14 World Precision Instruments...

Page 19: ...ecial or consequential arising from a failure of this product to operate in the manner desired by the user WPI shall not be liable for any damage to data or property that may be caused directly or ind...

Page 20: ...Tel 49 0 6031 67708 0 Fax 49 0 6031 67708 80 E mail wpide wpi europe com China Hong Kong WPI Shanghai Trading Co Ltd Rm 27A Liangfeng Building No 8 Dongfang Road Pudong New District Shanghai PR China...

Reviews: