pre-set volume and stop automatically. 

2)  You can also to set the water volume by time. The supply water time from 3 to 60secs. If 

the supply water time is over the 60secs, it will be stop automatically. 

3) 

It can be stopped when you press this button more time during water supplying.

 

6.3 Cuspidor Clean Water Function   

You can use the “Cuspidor clean” button to control the water to clean the Cuspidor. Press one 

time  to  open,  and  another  time  to  close.  The  volume  of  the  water  for  cleaning  could  be 

adjusted by throttle valve as followed picture. 

 

Pic 6-1

 

The time of cleaning the Cuspidor have 4 levels to be adjusted (1, 5, 30mins and Keep open), 

the process of setting as followed: 

Keep pressing “SET” button, every time to press ”Cuspidor clean” button will add the cleaning 

time a level,  and it  will be  cycling. The lamp near “SET” button  will flash corresponding time 

with the levels. Release ”SET” button to quit and save the setting.   

For example, if the default level is level 2 (5mins), keep pressing “SET” button, you can 

check followed form: 

Press ”Cuspidor 

Clean” button times 

Adding Level 

Lamp Flashing times 

Current Level 

1 time 

Increase 1 level 

3 times 

Level 3 (30mins) 

2 times 

Increase 2 level 

4 times 

Level 4 (Keep Open) 

3 times 

Increase 3 level 

1 times 

Level 1 (1min) 

4 times 

Increase 4 level 

2 times 

Level 2 (5mins) 

Release ”SET” button to quit and save the setting. 

CAUTION 

PLEASE NOT TO RELEASE “SET” DURING THE SETTING PROCESS. 

PLEASE  PRESS  “CUSPIDOR  CLEAN”  BUTTON  IN  7SECS  IN  CASE  TO 

ENTER ANOTHER LIMITED POSITION PROGRAM. 

 

Throttle Valve 

Summary of Contents for WODO Series

Page 1: ...WOSON Medical System Technical Publications WODO Series Dental Delivery Unit Operation Manual ASS0101 REV A Operation Documentation Copyright 2015 By Woson Medical Instrument Co Ltd...

Page 2: ...ulatory requirements of the following European Directive 93 42 EEC concerning medical devices Ningbo Jiangbei Woson Medical Instrument Co Ltd Address No 25 Lane 300 Jinshan Road Jiangbei District Ning...

Page 3: ...OR CHANGE Rev A 2015 07 15 First Issue Please verify that you are using the latest revision of this document Information pertaining to this document is maintained on manufactory If you need to know th...

Page 4: ...Italy Tel 39 02 84893641 Fax 39 02 84718594 This product complies with the regulatory requirement of the following Council Directive 93 42 EEC concerning medical devices The CE label affixed to the pr...

Page 5: ...fety Parts 11 2 4 Operation Risk 12 2 5 Protection Tool 12 Chapter 3 Description and Specification 13 3 1 Overview 13 3 2 Specification 14 Chapter 4 Receiving and Pre installation 16 4 1 Check the Pac...

Page 6: ...Receptacle Testing 30 6 5 Assistant Console Testing 31 6 6 Position Foot Control 31 6 7 X ray viewer 32 6 8 Foot Control 32 6 9 Chair position resetting program 32 6 10 Dental Unit Pre set I II III Po...

Page 7: ...res for operation and safety precautions 1 2 Usage Indications The chair mounted dental unit is intended to properly position a patient to perform dental procedures A dental operative unit is attached...

Page 8: ...ed on the label ATTENTION Pay attention to the high temperatures in the the dental unit exterior when exhausting system is running Protective Earth Indicates the protective earth grounding terminal CA...

Page 9: ...accordance with the instructions listed in this manual 2 2 1 Dental Chair The cushion do not less than 520mm and higher than 680mm Dental chair backrest angle swing should higher than 50 Dental chair...

Page 10: ...ttoon should retain the residue which the diameter is more than 2mm When the pressure is 200kPa the dirty on the cuspidor should be washed clean 2 3 6 3 way Syringe Well sealing performance open and c...

Page 11: ...e should be less than 70dB except the air compressor 2 2 15 The pressure system should be safe and reliable without leakage and explosion when in the normal condition 2 2 16 The parts which may genera...

Page 12: ...nment requirements of Group II climate environment and Group II mechanical environment GB T14710 93 The transportation test should follow the requirements of Chapter 3 GB T147710 93 The power supply v...

Page 13: ...l system for all the process of operation CAUTION MANUFACTURER IS NOT HELD RESPONSIBLE FOR ANY ARBITRARY DISASSEMBLY AMENDMENT WITH THE UNIT BY UNAUTHORIZED PERSON OR UNPROFESSIONAL TECHNITIAN 2 4 Ope...

Page 14: ...ure 3 Figure 3 Figure 3 1 The Structure of Dental Unit 1 The Structure of Dental Unit 1 The Structure of Dental Unit 1 The Structure of Dental Unit No Name Description 1 Foot Control Control water air...

Page 15: ...to provide air water and fog 16 Left Armrest Armrest on the left position of dental chair 17 Dental Chair The chair could be adjusted as up and down backrest front and back for different requirement 1...

Page 16: ...y an external source electric dental chair driven by DC motor Sign No input signal output signal Non movable common device splash proof foot switch Fuse Information Type Qty Position RT1 20 5 20 6 3A...

Page 17: ...340 2kg CAUTION WHEN YOU REMOVE THE PACKAGE PLEASE ENSURE THAT YOU DO NOT DAMAGE THE MACHINE INSIDE WHEN CUTTTING THE PLASTIC BAG BE CAREFULL NOT TO CUT THE TUBES INSIDE 4 2 Unpack the Accessories Ope...

Page 18: ...s Backup Gargle Hydrant with O Ring 1 pcs Cuspidor Filter 1 pcs Strong Suction Aspirator Filter 1 pcs Water Filer 1 pcs Screw M6 25 6 pcs Backup Water Connection Port 1 pcs Backup 4 Plastic Hoop 5 pcs...

Page 19: ...ore Setting Please ensure the place which has a separate and stable power source supply Please ensure the place which has the sewage draining tube to connect Please ensure the place which has the air...

Page 20: ...O NOT BEND THE POWER WIRE TO AVOID DAMAGE OF POWER WIRE DO NOT PUT ANY HEAVY STUFF ON THE POWER WIRE TO AVOID DAMGE OF POWER WIRE DO NOT USE OTHER POWER WIRE TO AVOID DAMAGE OF DENTAL UNIT DO NOT TRY...

Page 21: ...res 2mm 2 The power wire should be 2mm 2 3 wires with protecting bushing soft line leave power wires more than 400mm 5 2 Set Chair and Junction Box We suggest you to set the chair with a stationary ty...

Page 22: ...Ports Description Air and Water Tube Connection Check internal structure of ground box as Pic 5 2 put the water input tube draining tube power wire in the corresponding positions Water Input Port Air...

Page 23: ...in tube together with one port to drain Take out the draining collector and connect 3 tubes as above CAUTION PLEASE ENSURE THAT ALL THE CONNECTION PORT ARE SEALED WELL THE GREEN PU TUBE IS FOR WATER S...

Page 24: ...come a completed light arm with power wire from inside Decorative Ring and Cover Cross through Light Pole Completed Light Arm d Connect the power wire of light arm to the power wire connection port wh...

Page 25: ...rt Connect them with corresponding color wire 3 Color Wires Connected Completely c Insert the operation light into light arm and then fix limited screw Limited Screw Assembly Completely CAUTION PLEASE...

Page 26: ...y Lift the arm holder up till revolved then adjust the holder directly and press down for locking Lift the arm holder up and revolve Press down after adjusting 5 8 Balance Arm Assembly Lift the arm ho...

Page 27: ...control line to the delivery tray which inside the delivery tray CAUTION PLEASE ENSURE AT LEAST TWO PERSON TO ASSEMBLY THE BALANCE ARM TO AVOID ANY ACCIDENT 5 9 Handpiece and 3 Way Syringe Assembly C...

Page 28: ...2 Heater Button To control the water heating or not an indicator is near the button 3 Position 1 Return automatically to a set position 4 Position 2 Same function as above 5 Position 3 Same function a...

Page 29: ...the water heating state 6 2 Gargling Water Function The gargling water will be supplying when you pressing the Gargle water button stop automatically when reaching the pre set volume While pushing the...

Page 30: ...eep open the process of setting as followed Keep pressing SET button every time to press Cuspidor clean button will add the cleaning time a level and it will be cycling The lamp near SET button will f...

Page 31: ...dust air on the handpiece 1 Air brake switch 2 No 1 cooling water 3 No 2 cooling water 4 No 3 cooling water 5 No 4 cooling water 6 No 1 air Driving 7 No 2 air Driving 8 No 3 air Driving 9 No 4 air Dri...

Page 32: ...it Pic 6 4 Syringe Switch 6 5 Assistant Console Testing Take down the strong suction syringe to check if a strong suction strength from the port Take down the weak suction syringe to check if a weaker...

Page 33: ...for 8 seconds until indication light on 2 Press the button to the highest position as the chair can reach 3 Press the button to lower the chair position a little bit 4 Press the button to set backrest...

Page 34: ...e SET lamp blinking CAUTION DO NOT PRESS SET BUTTON MORE THAN 7SECS OTHERWISE IT ENTERS EXTREME POSITION PROGRAM 4 You can test by set another position and press 1 button if the chair could return the...

Page 35: ...ource Switch 2 Water Bottle 3 Change over Switch Water Source Switch to choose the source of bottle water or pipe water Bottle Air Switch To control air which into the bottle and press water supply ON...

Page 36: ...ersonnel Check the movement and operation Qualified service personnel As Required Sterilize Handpiece User Sterilize 3 way syringe User 7 1 Clean Filter 7 1 1 Clean Water Filter Water filter is placed...

Page 37: ...er The air filter is inside the control unit the yellow one as below picture showed When the suction syringe working it may suck inside much dust or impurity and left on the air filter It should be cl...

Page 38: ...of manometer then tight the value To release water as below Open the cover of junction unit Upward force the bottom of valve core pull the water out loose once the water is over B using the absorbent...

Page 39: ...sterilizer with 121 20mins 1 1bar The spittoon need to be cleaned and disinfected by alcohol each week Clean or replace the water filter core every half month Clean and sterilize the suction syringe...

Page 40: ...change the light 6 The movement range of the unit is very short Chair position not set properly Check Chapter 6 to reset the chair position 7 No mist cooling water from the handpiece The water cooling...

Page 41: ...rson The yearly maintenance and service should be executed by authorized personal The working life of this unit is 5 years 9 2 Safety Information Please confirm that the power plug is completely conne...

Page 42: ...r unplug the cord Discharge all the extra device such as air compressor scaler etc Drain the water from the tubes Discharge the operation light arm light and pack separately Use packing plastic film t...

Page 43: ...Dental Delivery Unit Operation Manual REV A Page 42 of 43 Appendix 1 Electric and Pipeline Diagram Electric Diagram S S S SE E E ET T T T...

Page 44: ...Dental Delivery Unit Operation Manual REV A Page 43 of 43 Pipeline Diagram...

Reviews: