Handpiece Lubricating System

 

Operation    Manual                                                                                                                    Page 3 of 12 

Dear Customer: 

Thank you for choosing Handpiece Lubricating System LUB909. The machine can make easy and safe 

cleaning and oiling on the Handpiece. Please read carefully the manual for maintenance. 

 Basic Features 

1. 

Ensure the cleansing and oiling of stomatology handpieces . 

2. 

Can serve three handpieces at the same time . 

3. 

The rotation gear of LUB909 can make effective cleaning and oiling for the handpiece in different angles. 

4. 

Press “Air” button after cleaning and oiling to remove residual oil.   

5. 

Short, long or extremely long mode can be selected according to the handpieces. 

6. 

The mist filter cotton can reduce the diffusion of the mist to the smallest extent. 

 

 Parameters 

1.Type:LUB909 

2. Rated Voltage: AC220V/50Hz, AC110V/60Hz         

3. Fuse

220V/240V

F1A       

250V

 

4. Air Pressure: 0.35

0.60Mpa (3.5

6.0kgf/cm2) (50

80psi) 

5. Capacity of Oil Container: 350mL 

6. Weight: 8 kg 

 

 Structure of Handpiece Lubricating System   

 

 

 

 

Accessories   

Pic 2

 

Pic 1

 

Panel

 

Oiling Cap

 

Door Cover

 

Oil Level Meter

 

Power Switch

 

Air Filter

 

Fuse Box

 

Power Input

 

Summary of Contents for LUB909

Page 1: ...Handpiece Lubricating System LUB909 Operation Manual NINGBO JIANGBEI WOSON MEDICAL INSTRUMENT CO LTD...

Page 2: ...age 2 of 12 Content Basic Features 3 Parameters 3 Structure of Handpiece Lubricating System 3 IV Installation Components 4 Operation Procedure 5 Disassembling and Installation of Front Door Cover 8 Ma...

Page 3: ...g and oiling for the handpiece in different angles 4 Press Air button after cleaning and oiling to remove residual oil 5 Short long or extremely long mode can be selected according to the handpieces 6...

Page 4: ...ing Piece 4 7 Oil Sorbent Cotton Pack 1 20 Slices 8 3 ways Connection Piece 1 IV Installation Components 1 Rated Fuse Follow the bellowing Table 2 to check whether the voltage is the same with rated f...

Page 5: ...Method on different kinds of Handpiece 1 Installation of Pneumatic Turbine Handpiece According to ISO9168 Remove drilling bit from the handpiece Open the door cover install the handpiece to the coupl...

Page 6: ...tor Power Indicator Mode Indicator S Short L Long E Extremely Long Coupling Head Selector Left Middle Right Air Key Start Pic 8 Pic 6 Remove the drill bit from the handpiece Open the door cover instal...

Page 7: ...Time for Cleaning Oiling Time for Air Blowing Short Mode Pneumatic Turbine Handpiece Handpiece with NAKANISHI NM series Handpiece with NAKANISHI SGM series About 25s About 60s Long Mode Corner Handle...

Page 8: ...When pushing the front door cover from right to left knot the moving pin into the right pin hole Maintenance 1 Replacement of Mist Filter Cotton Open the front door cover and take out the filter cott...

Page 9: ...ilter 1 Take out the air pipe from the quick coupling to the left and right as shown in Pic 14 and 15 2 Swing the screw with screwdriver and take off the air filter as shown in Pic 16 3 Make reversal...

Page 10: ...e As shown in Pic 18 to 21 Failure Recovery Check with the following table to determine whether similar circumstance can be found with the failure of the product If no contact your distributor for mai...

Page 11: ...tion Whether the handpiece is properly installed Inspection Maintenance Oil Leakage Whether the mode is selected or not Inspection Whether the drill bit hole of the corner handle handpiece is facing t...

Page 12: ...air filter should be clean and dry Make periodically wiping of the damp in the air compressor and air filter because the damp in the cavities will do harm to the machine when mixed with maintenance l...