7.4 CONTROLLER OTHER SETTINGS (ENGINEER MENU)
The following additional settings are only accessible to authorized service personnel!
NOTE For further details consult the separate manual (IM1026) or contact Worthington
Industries or your distributor.
Table 9
Engineer Menu's options of the controller
Service Call Indicator
Show/Hide User Calibration
Number of Thermocouple inputs
Aux Relay Function
Sim Fill Repeat
PC Software Type
Fill Timer, Start immediately or after Venting
Sim Fill Delay, Start immediately or after Venting
Analog Output ‘Polarity’
Set Memory Size
Set Baud rate for serial communications
7.5 CONTROLLER IN NORMAL OPERATION
During normal operation of the K SERIES Cryostorage System the controller display
shows the actual temperature of the thermocouple and the liquid level relative to the
sensors.
Filling takes place automatically to compensate for evaporative losses.
Filling may also be triggered by an external fill signal (Simultaneous Fill).
Once a simultaneous fill input signal has been accepted the controller will not respond
to another simultaneous fill input signal until 30 minutes have elapsed.
This is to avoid the possibility that two or more controllers could interact causing
repeated short fill cycles.
The 30 minute timer may be cleared by resetting the controller as described in section
7.7. This may be useful when testing for correct operation of a system connected for
simultaneous filling.