Description of the GAMMA FINDER® III
39 / 63
Test settings
Use this menu to define the measuring times for the stabil-
ity test as well as the nulleffect test. For the stability test, a
duration of 5 seconds to 180 seconds can be set. The longer
the measuring time, the more accurate the calculated value
will be.
For the nulleffect test, a duration of 5 seconds to 60 seconds
can be set.
Volume
Here you can adjust the volume of the acoustic output. You
can choose between "MUTE," "LOW," "MID," and "HIGH." The
selected value is confirmed on the screen and you return to
the OR screen.
Info
This menu displays important information about your
GAMMA FINDER® III.
This includes the built-in accumulator capacity in percent,
the version of the GAMMA FINDER® III, the device number,
the next maintenance date, and the current time stored in
the GAMMA FINDER® III.
Date/Time
This menu allows you to set the current date and time on
the device.
It is important that the correct date and time are stored in
the GAMMA FINDER® III or there may be deviations in the
stability test, for example.
5.6
Hygiene
Please refer to Chapter "Use of Sterile Sleeves/Drapes" [
for information on how to use the sterile sleeve (Article
No. 1031). The GAMMA FINDER® III is not approved for ster-
ilization. Non-compliance may diminish or eliminate the
functionality and safety of the GAMMA FINDER® III. W.O.M.
WORLD OF MEDICINE GmbH assumes no liability for any in-
cidents and damages.