Description of the GAMMA FINDER® III
23 / 63
WARNING!
Low or empty battery
Always use a fully charged device GAMMA FINDER® III. Us-
ing the device GAMMA FINDER® III with low or empty bat-
tery could cause stress to the operating team or harm the
patient.
5.5
Main Menu
Press the OK button to open the main menu. Use the
º
or
»
buttons to navigate through the menu.
5.5.1
Stability Test
The stability test serves to control the quality of the
GAMMA FINDER® III. The sensitivity of the GAMMA FINDER®
III and the background noise are checked.
Preparing the stability test (reference measurement)
OK
OK
OK
OK
OK
OK
OK
OK
Prior to performing a regular stability test for the first time
you need to register a source.
• To do this, start the GAMMA FINDER® III as described in
Chapter "Starting the GAMMA FINDER® III" [