background image

39

Î

 The teeth of the toothed belt underneath the slat-belt running surface have been pre-lubricated to 

reduce noise during operation. The toothed belt can be cleaned and lubricant can be applied to its edges 
once every six months or on an as needed basis. 

x

 Drive 

belt 

Î

 As with the running surface toothed belt, the edges of the drive belt can also be lightly greased if it 

"squeaks" audibly. 

x

 Inclination 

system 

Î

 This is also lubricated at the factory. If the system requires lubrication again, apply a small amount of 

lubricant to the lift mechanism shaft. 

Caution – Possible danger of injury or damage  

Please bear in mind that excess lubricant serves no purpose and only causes dirt and grease to 
accumulate. Apply only extremely small amounts of lubricant, and use only the original lubricant 
we have provided in the service kit. 

The Prescribed Recurring Test Period

The prescribed recurring test period (according to BGV A4 (VBG 4) Accident prevention directive) averages for 
treadmills 4 years (recommended is a test cycle of 1 year). This test may only be performed by an electrician or 
through an authorized person approved by the manufacturer.   

Electrical Testing According to VDE 0751

Grounding Conductor Resistance Measurement   

<0.3 Ohm   

Isolations Resistance Measurement  

 

>2 MOhm   

Backup Electrical Measurement 

 

 

<1 mA  

These measurements should be performed with approved gauges according to VDE 0751. A MTK “Technical 
Measurement Control” according to § 11 is not prescribed by Woodway. If fluctuations of the display occur, the 
measuring unit should be cleaned.    

If a Technical Measurement Control is requested, a form from WOODWAY is available or it can be performed by 
a WOODWAY approved Service Technician.  The means and breadth of the Technical Measurement Control will 
be defined by WOODWAY.   

Summary of Contents for LokoStation PPS

Page 1: ...Operating Instruction LokoStation WOODWAY GmbH Steinackerstr 20 79576 Weil am Rhein Germany Phone 49 0 7621 940 999 0 Internet www woodway de Fax 49 0 7621 940 999 40 e mail info woodway de Rev 01 06...

Page 2: ...and up to date as possible Insofar as legal limits allow we reject any liability for damages resulting from use of this operation manual At most we accept liability for intent and culpable negligence...

Page 3: ...the handrail mounted keypad 13 6 3 WOODWAY User System WUS 15 6 3 1 Switching the treadmill ON 15 6 2 3 Manual Use of the Handrail mounted keypad 18 6 2 4 Start Program 19 6 2 5 Edit Program 21 6 2 6...

Page 4: ...he values listed on the nameplate of the slat belt treadmill Any modification of the input voltage should be performed only by WOODWAY or by their duly authorized service agent The treadmill must be c...

Page 5: ...ust be performed only by the manufacturer or their duly authorized service agents never by the operator We are not liable for damages resulting from unauthorized service or from improper maintenance p...

Page 6: ...and it will also protect the floor covering from undue wear Due to the weight of the LokoStation with treadmill ca 1 764lbs the ceiling or the floor should have the necessary support capability or th...

Page 7: ...position it underneath the treadmill and then move the treadmill to the desired new location If the treadmill must be disassembled the panels and rails can be removed for ease of transportation refer...

Page 8: ...on and designations 1 Emergency Stop pull wire magnet 2 Handrail mounted keypad 3 Lifting bar insertion into the hole for which it was designed 4 RS 232 USB interface 5 ON OFF main power switch Power...

Page 9: ...tion 2 One Arm Construction 3 Body Static Dynamic Body Weight Support 4 Rope Length Adjustment Variations possible 5 Central Winch Static Mode 6 Crank Dynamic Body Weight Support 7 Security Spikes for...

Page 10: ...tion of the Hips x x x x Dynamic Static Body Weight Support x x x x Symmetric Asymmetric Body Weight Support x x x x Variable Deviation Points x x x x Wheel Chair Ramp incl Handrail left x x x x 2 adj...

Page 11: ...n against electrical shock equipment for external and internal use on patients however not for direct use on the heart Protective conductor connection grounded for DIN 40011 Attention Before working o...

Page 12: ...area the 230 V power supply is interrupted and an Emergency Stop is initiated The magnet must be positioned on the red field with the yellow ring When the magnet is not correctly positioned the tread...

Page 13: ...of the Emergency Stop systems 6 2 Data Monitor The WOODWAY data display module is used to show the treadmill s current control parameters and modes In addition the Standby Button and the Emergency Sto...

Page 14: ...down press the blue button Speed will diminish gradually You can either press this button several times in succession or hold it down continuously until the desired speed has been reached To stop more...

Page 15: ...ink once per second The upper LED to the right of the display indicates this condition When the length of a training session exceeds one hour the display changes to hours and minutes At that point the...

Page 16: ...rol panel In addition the treadmill can also be operated i e via speed and inclination settings from the keypad on the handrail 1 Touch screen 2 Standby button 3 Emergency Stop pull wire magnet 4 Cont...

Page 17: ...red up the treadmill with the main ON OFF switch located under the right side as viewed from the running direction activate the treadmill by pressing the Standby button 2 To stop the treadmill please...

Page 18: ...n hh mm ss x Heart rate in bpm beats per minute If the Polar transmission chest strap is used the pulse rate in beats per minute bpm is displayed Caution Possible danger of injury or damage Although t...

Page 19: ...l operation mode the direction of travel automatically resets to forward a Inclination adjustment To activate the inclination for running uphill or downhill the latter being an optional feature press...

Page 20: ...or up and down for marking the desired program Program names are displayed together with indicators showing whether they are package or custom programs SELECT loads the program of your choice i e the...

Page 21: ...for diagnostic or medical purposes For diagnostic or medical purposes use only ECG instruments specifically designed and approved for that purpose x Power in watts or energy value in kcal To select di...

Page 22: ...the button EDIT PROGRAM The next screen then allows you to select the following operations SELECT PROGRAM This switches you to the Treadmill programs menu where you can select a treadmill program as...

Page 23: ...le shown above press the button four more times until the display shows Step 5 of 21 SPEED etc Select the parameter that you would like to change Modify the given parameter by entering the new value C...

Page 24: ...s that you have modified DELETE deletes the currently displayed program step INSERT inserts a new program step You must enter the parameters for this program step separately as described above CANCEL...

Page 25: ...llowing mask Press the SAVE button only at the very end after you have completed all entries In the following mask you will be asked to confirm that you really want to save the program SAVE stores the...

Page 26: ...e pulse measurement system is only used as a general indicator The treadmill pulse measurement is not intended for diagnostic or medical purposes For diagnostic or medical purposes use only ECG instru...

Page 27: ...SE SETPOINT Use the number keypad on the touch screen to enter this value see above LIMIT SETTINGS This allows you to enter limits for speed and inclination The treadmill generally adjusts itself to a...

Page 28: ...llowing screen appears Press the corresponding button to adjust the desired value The programmed value is displayed in this mask e g speed in km h Press the END button to confirm acceptance of the new...

Page 29: ...erformance parameters of the treadmill The contents of this display screen are only for your information these parameters cannot be modified Press the Arrow button to access further technical informat...

Page 30: ...f the Bar Rail are adjustable Gas Pressure Blocking Springs which allow combined with an easy to use clamping fixture for locking the Bar a uniquely simple handling Figure 1 Push button for locking th...

Page 31: ...sired height This can be done in both directions up down The Rail moves up almost automatically light pull is needed downwards the Rail has to be pushed against the resistance of the Gas Spring Fig 2...

Page 32: ...e COG throughout the whole gait In every single phase of patient movement it is guaranteed that the chosen weight supports the patient As a precaution check the patient s ability to stand alone before...

Page 33: ...e the following tool for adjusting the shoulder width Figure 3 Setting the Deviation Points Take the adjustable knob shown in the picture pull it down for release and shift it to the left or right sid...

Page 34: ...spikes should be pushed in and visible over the weight change plates As a result dynamic weight support will be prevented Figure 5 Central Winch The elevation ropes will be pulled up by turning the ce...

Page 35: ...counter clockwise direction At this point the Weight Support System is in the static position i e the security spikes are pushed in Figure 6 Setting the Dynamic Body Weight Support With the crank one...

Page 36: ...inside the drive link sink to the lowest position As a result one can reset the security spikes and the weight support system is back in the static weight position 6 5 4 Rope Length Adjustment The rop...

Page 37: ...aching the patient the 2 carabiners should be brought to a middle position default position see figures A B A B Pulling down the rope adjustment will lift up the left body half of the patient see figu...

Page 38: ...By loosen the rope adjustment the left body half will be lowered see figure E F E F To guarantee a safe fixation of the rope the a self locking rope By pulling the rope it gets automatically fixed...

Page 39: ...d off and the main power cable must be unplugged 7 2 Treadmills in the Medical Market Weekly x Wipe the handrails control panel and exterior panels with a damp cloth x Check the power cable for cracks...

Page 40: ...we have provided in the service kit The Prescribed Recurring Test Period The prescribed recurring test period according to BGV A4 VBG 4 Accident prevention directive averages for treadmills 4 years re...

Page 41: ...ion x All security relevant screws Every six 6 Months x Examine the T Tracks Every twelve 12 Months x Exchange replace all ropes x Exchange replace all pulleys x Examine the entire system for potentia...

Page 42: ...he following points x Is the treadmill switched on main switch x Is the treadmill properly connected to an electrical socket power cable correctly inserted into the treadmill and the electrical socket...