Maintenance and Cleaning
10/2015-v3.4en
55
9.4.2
LokoStation ELVETA
Defect
Cause
Solution
Remote control does not function
Remote control or extension cord
is damaged
Replace remote control or exten-
sion cord
ELVETA does not function, only
emergency lowering
Electronics defective
ELVETA must be checked by the
manufacturer or a service partner
Fuse defective
Motor has been overloaded
The maximum load of 160 kg has
been exceeded
Power supply is broken
Check the output voltage of the
power supply.
Lifting/lowering does not function
Check the following points:
-
Is the belt (rope) under
tension when the function
switch is pressed? (7.5.2)
-
Does the green LED light on
the drive (cabinet)?
-
Is the main switch on at the
drive? (7.5.4)
-
Is the fuse damaged? (7.5.4)
-
Is the power adapter func-
tioning?
-
Does the emergency lowering
function? (7.5.5)
-
If the above points are in
order, please contact your
service partner
During lifting/lowering the patient
drops down to a few centimeters
Button for emergency lowering
was pressed without tension on
the belt - belt is tangled in the
motor
Unwind the belt completely and
rewind it under tension
Noises occur when the emergen-
cy lowering is activated: Hum-
ming
Inadequate power supply
Either the ELVETA hadn't been
used for a time and the batteries
have lost capacity, OR the
charger is defective and the
batteries have been discharged
ELVETA must be checked by the
manufacturer or a service partner