14
2. It should not be mixed with dangerous goods during transportation.
3. Avoid exposure to sun or rain or snow during transportation.
The cleaning and disinfection of main unit are as follows.
• Before each use, wipe the surface of the connector of cable and
cable with a soft cloth or paper towel soaked in 75% medical alcohol.
Repeat the wipe for at least 3 times.
• Before each use, please let the Ultrasonic scaler work under
irrigation mode for 20-30s, then install the handpiece.
• After each use, please let the Ultrasonic scaler work under irrigation
mode for 20-30s, then remove the handpiece.
• After each use, wipe the surface of the connector of cable and cable
with a soft cloth soaked in clean water (distilled or deionized water) or a
clean disposable wipe. Repeat the wipe for at least 3 times.
5. Precaution
5.1 usage notice
5.1.1 Keep the scaler clean before and after operation.
5.1.2 The scaling tip, wrench and handpiece must be sterilized before each
treatment.
5.1.3 Don't screw the scaling tip when stepping on the foot switch.
5.1.4 The scaling tip must be fastened. There must be fine spray coming out
from the tip when operating.
5.1.5 Change a new one when the tip is damaged or worn excessively.
5.1.6 Don't twist or rub the tip.
5.1.7 While scaler working, the heat of scaling tip will become higher if there
is no water flowing out, please keep the water flow smoothly.
5.1.8 Don't use impure water source, and be sure not to use normal brine
instead of pure water source.
5.1.9 If use the water source without hydraulic pressure, the water surface
should be one meter higher than the head of the patient.
5.1.10 Don't knock or rub the handpiece.
5.1.11 After each clinical operation with clinical liquid, change a bottle with
purified water, turn the water supply to max, make the machine work with auto-
water supply for 30 seconds in order to keep the water way and spare parts clean
and durable.
5.1.12 Please put the power plug into the socket easy to pull out, to make sure
it can be pull out in emergency.
5.1.13 When using the equipment, please keep the water get through
smoothly, otherwise patient's tooth surface would be injured by overheat in the
handpiece.