Woodpecker U600 LED Instruction Manual Download Page 1

U600 LED  ULTRASONIC  SCALER

INSTRUCTION MANUAL

Summary of Contents for U600 LED

Page 1: ...U600 LED ULTRASONIC SCALER INSTRUCTION MANUAL ...

Page 2: ...nance 4 4 Cleaning Disinfection and Sterilization 8 5 Precaution 14 6 After service 16 7 Symbol instruction 16 8 Environmental protection 16 9 Manufacturer s right 17 10 European authorized representative 17 11 EMC Declaration of conformity 17 12 Statement 20 TABLE OF OPERATING POWER OF THE TIPS 21 ...

Page 3: ...scaler U600 LED is composed of electrocircuit water way and ultrasonic transducer 1 2 3 Scope of application Ultrasonic scaler U600 LED is used for the dental calculus elimination and root canal treatment 1 3 The main technical specifications 1 3 1 Technical specifications of ultrasonic scaler a Main unit input 220 240V 50Hz 60Hz 150mA b Output primary tip vibration excursion 90μm c Output half ex...

Page 4: ...xture with air oxygen or nitrous oxide 1 4 Instruction of the main components Instruction and component sketch map 1 4 1 Front of the main unit sketch map Picture 1 1 4 2 Back of the main unit sketch map Picture 2 1 4 3 Instruction of using wrench ...

Page 5: ... Open the packing box make sure that all the parts and accessories are complete according to the packing list take the main unit out of the box and put it on the the stable plane facing to the operator 2 1 2 Turn the water control knob to the max according to the picture 1 ...

Page 6: ...ccording to your need usually adjust to the middle grade and adjust the vibrating during the clinical treatment according to the patient s sensitivity and the rigidity of the tartar 2 1 9 Step on the foot switch the tip begins to vibrate and the LED lamp on the top of the handpiece shines Release the foot switch the LED lamp keep shining for 10 seconds 2 1 10 Water volume adjustment Step on the fo...

Page 7: ...act Contact our dealers or us Something wrong with the handpiece Send the handpiece to our company to repair Something wrong with the cable Contact our dealers or us The scaling tip vibrates but there is no spray when stepping on the foot switch The water control knob is not on Turn on the water contrl knob note 1 The vibration of the tip becomes weak The tip hasn t been screwed on to the handpiec...

Page 8: ...er pipe is jammed Clean water pipe with multi function syringe note2 The u file doesn t vibrate The screw is loose Tighten it Endochuck is damaged Change a new one There is noise coming from the endochuck The screw is loose Tighten it LED light don t work Poor contact Contact tightly Something wrong with LED light Change a new one There is no water coming out from the handpiece automatic water sup...

Page 9: ... 3 Cleaning mode The liquid pipeline of the machine must be rinsed after each use of the clinical liquid to reduce the crystal accumulation and the bacteria in the liquid pipe Operation 1 Put distilled water or mineral substance into water tank 2 Press the automatically water supply button and outside water button at the same time 1s to start the Cleaning mode after buzzer beeps The button of auto...

Page 10: ... is 1000 times 4 1 Initial processing 4 1 1 Processing principles It is only possible to carry out effective sterilization after the completion of effective cleaning and disinfection Please ensure that as part of your responsibility for the sterility of products during use only sufficiently validated equipment and product specific procedures are used for cleaning disinfection and sterilization and...

Page 11: ...cable front thread horn nipple seal ring light pipe and LED lamp if any until the dirt on surface is not visible Then use soft cloth to dry the product and accessories and put them into a clean tray The cleaning agent can be pure water distilled water or deionized water Disassembling steps 4 3 Cleaning The cleaning should be performed no later than 24 hours after the operation The cleaning can be ...

Page 12: ...ture disinfection function The temperature does not exceed 134 C and the disinfection under the temperature cannot exceed 20 minutes The disinfection cycle is in accordance with the disinfection cycle in EN ISO 15883 Cleaning and disinfecting steps by using Washer disinfector 1 Carefully place the product into the disinfection basket Fixation of product is needed only when the product is removable...

Page 13: ...mperature is 93 C the time is 2 5 min and A0 3000 e Only distilled or deionized water with a small amount of microorganisms 10 cfu ml can be used for all rinsing steps For example pure water that is in accordance with the European Pharmacopoeia or the United States Pharmacopoeia f After cleaning the chemical residue should be less than 10mg L g The air used for drying must be filtered by HEPA h Re...

Page 14: ...or replacement must be cleaned disinfected and dried 4 6 4 If the service time number of times of the product reaches the specified service life number of times please replace it in time 4 7 Packaging Install the disinfected and dried product and quickly package it in a medical sterilization bag or special holder sterile box Notes a The package used conforms to ISO 11607 b It can withstand high te...

Page 15: ...lity for the procedures that have not been recommended If you use the sterilization procedures that have not been recommended please adhere to related effective standards and verify the suitability and effectiveness Fractional pre vacuum procedure steam sterilization with repetitive pre vacuum The procedure used here is to perform steam sterilization through three pre vacuums 4 9 Storage 1 Store i...

Page 16: ...he scaling tip when stepping on the foot switch 5 1 4 The scaling tip must be fastened There must be fine spray coming out from the tip when operating 5 1 5 Change a new one when the tip is damaged or worn excessively 5 1 6 Don t twist or rub the tip 5 1 7 While scaler working the heat of scaling tip will become higher if there is no water flowing out please keep the water flow smoothly 5 1 8 Don ...

Page 17: ... be cautious to use the equipment 5 3 Storage and maintenance 5 3 1 The equipment should be handled carefully and lightly Be sure that it is far from the vibration and is installed or kept in a cool dry and ventilated place 5 3 2 Don t store the machine together with the articles that are combustible poisonous caustic or explosive 5 3 3 This equipment should be stored in a room where the relative ...

Page 18: ...ment Type BF applied part Foot switch interface Used indoor only Adjustment for the water flow Sterilizable up to the temperature specified Mode of outside water system Mode of auto water system Water entrance pressure Follow Instructions for Use CE marked product Earth ground Adjustment for the water flow in Mode of auto water system Appliance compliance WEEE directive Atmospheric pressure for st...

Page 19: ...e technique fittings the instruction manual and the content of the original packing list at any time without notice If there are some differences between blueprint and real equipment take the real equipment as the norm 10 European authorized representative 11 EMC Declaration of conformity ...

Page 20: ...18 ...

Page 21: ...19 ...

Page 22: ...ce in high electromagnetic environment 12 Statement All rights of modifying the product are reserved to the manufacturer without further notice The pictures are only for reference The final interpretation rights belong to GUILIN WOODPECKER MEDICAL INSTRUMENT CO LTD The industrial design inner structure etc have claimed for several patents by WOODPECKER any copy or fake product must take legal resp...

Page 23: ... 3 E G 7 1 10 G E5 1 6 E G 8 1 10 G E5D 1 3 E G 9 1 10 G E8 1 10 E G 10 1 10 G E9 1 10 E G 11 1 10 G E10 1 6 E E10D 1 6 E Periodontics E11 1 6 E Tip Model Power E11D 1 6 E P1 1 10 P E14 1 3 E P2L 1 3 P E15 1 3 E P2LD 1 2 P P2R 1 3 P Cavity Preparation P2RD 1 2 P Tip Model Power P3 1 6 P SB1 1 10 P P3D 1 6 P SB2 1 10 P P4 1 6 P SB3 1 10 P P4D 1 6 P SBL 1 10 P SBR 1 10 P ...

Page 24: ...ZMN SM 059 V1 9 20201124 ...