25
4.8 Canal measurement operation
When using as alone apex locator mode.
We suggest put the motor handpiece on
the charging base to get better visual
angle.
Press Setting button “P” once during
standby state, press Adjusting button
“+”/“-” to select EAL Operation mode,
then press Main button to confirm. (See
chapter 3.3 Terms and definition to get
the explanations of Operation modes.)
The measuring wire must be connecting
with motor handpiece by USB socket,
white socket connects with patient’s lip
by lip hook, and black socket connect
with file clip.
The canal length indicator bar will show
on the screen(more information about
canal length indicator bar, please see
chapter 3. 2 Screen display).
The file clip must hold the file correctly.
Push the button on the file clip with
your thumb in the direction shown by
the arrow. Clip the holder onto the metal
upper part of the file and then release
the button.