20
Forward Angle: only activating in REC and
ATR operation mode.
Reverse Angle: only activating in REC operation
mode.
F: Forward Angle
R: Reverse Angle
Press Adjusting button “+”/“-” to change angle,
adjustable every 10 degrees.
It is suggested that the difference between the
forward angle and reverse angle should be
greater than or equal to 120 degrees, otherwise,
root canals cannot be prepared effectively.
Forward Angle<Reverse Angle, such as F: 30º/
R: 150º, effective cutting angle is Reverse Angle,
it is suitable for used the reciprocating files likes
DENTSPLY WAVEONE or WOODPECKER
W3-ONE.
Forward Angle>Reverse Angle, such as F: 180º/
R: 30º, effective cutting angle is Forward Angle,
it is suitable for used the reciprocating files likes
SENDONELINE S1.
Remarks: only 120°~340° forward angles are
available in ATR mode.
4.4 Preset program selection
For convenience, we preset some common file
system.
Press Adjusting button “+”/“-” to switch to
preset program(M0-M9, preset program 1-5 ),
the interface will show as left.
Long press Setting button “P” to entry preset
program during standby state, the interface will
show as left.
Press Adjusting button “+”/“-” to select file
system.