Page 1: ......
Page 2: ...tion instruction 16 5 Troubleshooting 30 6 Cleaning Disinfection and Sterilization 31 7 Storage maintenance and transportation 34 8 Environmental protection 35 9 After service 35 10 European authorize...
Page 3: ...motor with root canal measurement capability It can be used as a endo motor for preparation and enlargement of root canals or device for measuring canal length It can be used to enlarge the canals whi...
Page 4: ...2...
Page 5: ...3...
Page 6: ...onment 1 7 4 This device requires special precautions regarding electromagnetic compatibility EMC and must be in strict accordance with the EMC information for installation and use Do not use this equ...
Page 7: ...to be used for some time 1 7 17 Wireless charging will generate heat and the surface temperature of charging base and motor handpiece will rise It is recommended that the time of contacting motor hand...
Page 8: ...Speed rang 100rpm 1200rpm 1 9 5 Wireless charging only for Ai Motor Frequency range 112 205KHz Maximum RF output power of the product 9 46dBuA m 3m 1 10 Environment parameters 1 10 1 Environment temp...
Page 9: ...d Standby 2 2 1 1 EAL Mode This mode is for canal measurement The motor handpiece does not run in this mode 2 2 1 2 CW Mode The motor handpiece rotates forward 360 clockwise direction Used rotaty file...
Page 10: ...ing files likes DENTSPLY WAVEONE or WOODPECKER W3 ONE Forward Angle Reverse Angle such as F 180 R 30 effective cutting angle is Forward Angle it is suitable for used the reciprocating files likes SEND...
Page 11: ...s passed the apex foramen The digital number 00 indicate that the file has reached the apex foramen Subtract 0 5 1mm from the measured file length as the working length These numbers are used to estim...
Page 12: ...urely into place The contra angle rotates 360 so that the OLED display can always be viewed easily 2 4 2 Removal Pull out the contra angle horizontally when the motor handpiece does not run Warnings a...
Page 13: ...r plugging the file into contra angle let go the hand on push cover to assure that the file cannot be taken out Be careful when inserting files to avoid injury to fingers Inserting and removing files...
Page 14: ...ed if the canal measurement function will not be used Connect the measuring wire to the motor handpiece Line up the measuring wire plug with the notch on the back of the motor and push it all the way...
Page 15: ...by sate d Parameter setting Press Setting button P till target parameters press Adjusting button to change then press Main button or wait 5 seconds to confirm e Preset program selection Long press Set...
Page 16: ...rface a Set speed b Set torque c Real time torque d Torque display scale Canal measurement mode interface a Apical reference point flash bar b EAL Electronic apex locator Canal measurement state inter...
Page 17: ...tive torque reverse Up to setting torque the motor will move with reciprocating ATR mode when torque reduce to normal value the motor will clockwise rotate Forward Angle Activating in REC and ATR oper...
Page 18: ...and CCW modes the torque value Torque Limit that triggers reverse rotation For ATR mode the torque value Trigger Torque that triggers ATR action 4 Operation instruction 4 1 Power on and power off 4 1...
Page 19: ...rs can be changed 4 3 Parameter setting Before starting of motor handpiece please check the operation mode is correct All the parameters must be set according to files make sure all the parameters are...
Page 20: ...EC mode the torque of 2 0Ncm 5 0Ncm are available Actions that happen automatically when the file tip reaches the point inside the canal determined by the Flash Bar setting Benefit from integration of...
Page 21: ...ange OFF Motor does not stop when file is taken out the canal The Main button is used to start and stop the motor handpiece ON Motor stops automatically This is the reference point where various apica...
Page 22: ...is Reverse Angle it is suitable for used the reciprocating files likes DENTSPLY WAVEONE or WOODPECKER W3 ONE Forward Angle Reverse Angle such as F 180 R 30 effective cutting angle is Forward Angle it...
Page 23: ...he preset program default setting is not recommended otherwise has risk of file separate 4 5 Handpiece functions setting With the motor handpiece turned off hold down the Setting button P and press Ma...
Page 24: ...be calibrated before use Press Setting button P again the Beeper Volume can be change press Adjusting button to adjust then press to Main button to confirm The Beeper Volume can be set from 0 3 Vol 0...
Page 25: ...ery capacity the low battery capacity is insufficient to support the motor handpiece to reach the limit torque value that is the auto reverse function will not work properly Please charge it in time 5...
Page 26: ...ength indicator bar please see chapter 3 2 Screen display Setting parameters of automatic functions as needed such as Apical Action Auto Start etc more information about automatic functions please see...
Page 27: ...ations of Operation modes The measuring wire must be connecting with motor handpiece by USB socket white socket connects with patient s lip by lip hook and black socket connect with file clip The cana...
Page 28: ...suring wire or file clip should be replace Root canals not suitable for canal measurement Accurate measurement cannot be obtained if the root canal conditions shown below Root canal with a large apica...
Page 29: ...n is overflowing from the canal opening In this case clean the canal and its opening It is important to get rid of any solution overflowing the opening Broken crown If the crown is broken and a sectio...
Page 30: ...e opening at the top of the crown so that the file will not touch the metal prosthesis before taking a measurement Extremely dry canal If the canal is extremely dry the meter may not move until it is...
Page 31: ...s that the battery capacity is enough and there is no need to charge After charging please unplug the power adapter 4 10 Replacing Battery Replace the battery if it seems to be running out of power so...
Page 32: ...ooting Failure Possible cause Solutions The motor handpiece does not rotate Chose EAL mode EAL mode is only for canal measurement Changing to CW CCW REC or ATR mode There is continuous beep sounds aft...
Page 33: ...king FDA and Health Canada approval and in accordance with the DFU of the disinfecting solution manufacturer 6 2 3 Do not place the contra angle in a disinfectant solution or in an ultrasonic bath 6 2...
Page 34: ...with washer disinfector Put the accessories contra angle lip hook file clip touch probe protective silicon cover into the washer disinfector Ao value 3000 or at least 5 min at 90 C 194 F Avoid any co...
Page 35: ...ouch given by the manufacturer to determine the shelf life Use packaging which is resistant to a temperature up to 141 C 286 F and in accordance with EN ISO 11607 5 Sterilization Steam sterilization a...
Page 36: ...shorten the life of electronic components damage battery reshape or melt some plastic 7 1 3 Avoid the storage in a too cold condition Otherwise when the temperature of the equipment increases to a no...
Page 37: ...local laws 9 After service From the date this equipment has been sold based on the warranty card we will repair this equipment free of charge if there are quality problems Please refer to the warrant...
Page 38: ...not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electromagnetic environment Technical Description Concerning Electromagnet...
Page 39: ...ustomer or the user of the model Ai Motor MotoPex should assure that It is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electros...
Page 40: ...Motor MotoPex be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field lEC 61000 4 8 30A m 30A m Power frequency magnetic fields should be at levels charac...
Page 41: ...MHz to 800 MHz d 2 3 P1 2 800 MHz to 2 7 GHz where P is the maximum output power rating of the transmitter In watts W according to the transmitter manufacturer and d Is the recommended separation dist...
Page 42: ...s than 3V m Table 4 Recommended separation distances between portable and mobile RF communications equipment and the model Ai Motor MotoPex Recommended separation distances between portable and mobile...
Page 43: ...the transmitter where P is the maximum output power rating of the transmitter in watts W accordable to the transmitter manufacturer NOTE I At 80 MHz and 800 MHz the separation distance for the higher...
Page 44: ...ZMN WI 09 475 V1 4 20190703...