Read these instructions carefully to familiarize yourself with the product. Please retain
these operating instruction for future reference.

Use

Safety Bunsen Burner for Microbiology, Cell Biology, Dental Labs, etc. Ideal for
heating and flame sterilizing

         WARNING: DO NOT LEAVE THE ACTIVATED SAFETY

           BUNSEN BURNER UNATTENDED!

Safety Precautions

On unpacking the unit, check for possible transportation damages. Do not
operate the unit if damages are visible.

After use or for any longer period of time without attendance, turn the main gas
supply off. To depressurize gas hose keep the main gas supply off and activate
the burner again until the flame extinguishes in order to burn the residual gas.
Afterwards turn off the gas burner at the function knob (1).

Pay attention to your relevant rules for using liquid gas.

Only use DVGW safety tubings with thread or tubing connectors. Check the
condition of the tube/hose frequently. Depending upon type of tube/hose, hose
clamps are required.

• All gas connections must be adequately tightened with two wrenches. Ensure

gas proofness with a suitable test fluid / equipment. DO NOT seal the swivel nut
(L2) with Teflon tape etc.

Keep hands or other parts of the body away from the burner orifice (5a).

Do not operate the unit near flammable liquids or hazardous materials.

Unattended  operation of the unit is not permissible.

Always work in a well-ventilated area.

Note that the burner tube and burner orifice (5, 5a) remains hot after the flame
has been extinguished. Do not touch. Can cause burns.

Allow sufficient time for burner tube (5) to cool down prior to
cleaning, desinfecting, servicing or transport. Ensure that the
unit and the gas supply are turned off.

Allow sufficient time for burner tube (5) to cool down prior to disassembling.

Operate the unit with assembled burner tube (5) only.

After cleaning the burner tube (5) allow sufficient time to dry before assembling
again.

Keep substances away from the flame orifice (5a).

Summary of Contents for Flame 100

Page 1: ...1 EN Sets new standards in the Bunsen burner class Instruction manual DIN30665 1...

Page 2: ...foot pedal 7 Power connector for 9 V DC 8 Gas inlet R 1 4 L gas adapter lefthandthread 9 Holding device for inoculationloopholder 10 Electrodeholder 10a Screw for Electrode holder 11 Monitorelectrode...

Page 3: ...2 4 Foot pedal connection External IR motion sensor 7 3 Operation 7 3 1 On Off switch 7 3 2 Application programs 7 3 3 Flame regulation 7 3 4 Switch off End of work 7 4 Safety Symbols Safety Control S...

Page 4: ...rs Check the conditionofthetube hosefrequently Dependingupontypeoftube hose hose clamps are required All gas connections must be adequately tightened with two wrenches Ensure gasproofnesswithasuitable...

Page 5: ...rner tube 5 by turning counter clockwise and unscrew it completely Loosen the nozzle N 55 14 for natural gas with the wrench L1 supplied by turning counter clockwise unscrew completely and remove Ther...

Page 6: ...or other gas approved pressure regulator 50mbar must be used for liquid gas Pay attention to your relevant rules for using liquid gas Several gas cartridge adapter are optionally available 7 8 6 2 3...

Page 7: ...Foot pedal connection External IR motion sensor Insert the connection cable of the foot pedal or the external IR motion sensor accessory into the socket 6 at the back of the unit The foot pedal extern...

Page 8: ...off and the unit indicates a malfunction see paragraph 5 5 5 Error displays Ignition failure Status LED 4 blinks red 2x This signal appears and indicates a malfunction if the flame fails to ignite af...

Page 9: ...r wash it in a dishwasher To remove the burner tube proceed as follows Loosen the screw for the electrode holder 10a with the flat head screwdriver supplied L3 Slightly tilt the electrode holder 10 ba...

Page 10: ...against any manufacture defects in material and workmanship The WLD TEC warranty guarantees all gas burners under normal usage conditions and does not cover any damages as a direct result of user misu...

Page 11: ...ollowing standards togetherwiththenormativedocumentA2mainsconnection 1 ElectromagneticCompatibilityDirective 1 1 EN 61326 1 2006 Electricalequipmentformeasurement controland laboratoryuse EMCrequireme...

Page 12: ...bar liquid gas 20 50 mbar Connected load 43 g h liquid gas 53 l h natural gas E 49 l h natural gas LL Continuous cartridge operation CV 360 65 min CG 1750 210 min approx C 206 230 min CP 250 305 min C...

Page 13: ...tube 5 Especially remove substances in the marked areas at the electrodes If there are contaminants in this area the flame cannot encircle the electrodes correctly Clean this areas with a brush The b...

Page 14: ...ee paragraph 5 Status LED blinks red 4x Overtemperature see paragraph 5 Status LED flashes amber BHC activ see paragraph 4 Service address WLD TEC GmbH Production Service Halle Kasseler Str 49 37318 A...

Page 15: ...G ttingen Telefon 49 0 551 793789 Fax 49 0 551 793707 Internet http www wld tec com ProductionundService Halle KasselerStra e49 D 37318Arenshausen Telefon 49 0 36081 68940 Fax 49 0 36081 68942 Email...