Electromagnetic compatibility 

The product complies with the EMC-standard according to IEC60601-1-2, but the mobile communicat

ion devices’ emissions can disturb the product’s 

functioning. For the proper functioning of the device please follow the instructions of the user manual.  
 

Electromagnetic disturbance emission 

T

HE 

WIWE

 MOBILE AND HEART DIAGNOSTIC DEVICE WAS MADE TO BE USED IN AN ELECTROMAGNETIC ENVIRONMENT WITH THE CONDITIONS BELOW

.

 

 

T

HE USER NEEDS TO ENSURE

,

 THAT THE DEVICE IS FUNCTIONING WITH THESE CONDITIONS

.

 

 

Disutrbance emission examination 

Compliance 

Electromagnetic environment 

– guideline 

RF emissions,  
MSZ EN 60601-1-2:2008, MSZ EN 55011:2010 

Class B 

Approved: 30-

1000MHz,  

 

The WIWE device’s consumption is lower, than 75 W. Its disturbance 
emission is negligible. so it won’t cause any disturbances in the 
functioning of the device. 

 WIWE mobile diagnostic deivce is suitable to use in any kind of 

establishment, including houses and establishments directly connected 
to the low voltage network supplying the houses. 

 The WIWE mobile heart diagnostic device is not suitable to be connected 

with other devices. 

RF emissions, supply gateway 
MSZ EN 60601-1-2:2008, MSZ EN 55011:2010 

Class B 

Approved: 0,15-30MHz

 

Overtone emissions 
MSZ EN 60601-1-2:2008, MSZ EN 61000-3-2 : 
100Hz-2kHz 

[

non applicable]

 

Changes in voltage /flicker emissions 
MSZ EN 60601-1-2:2008, MSZ EN 61000-3-3:2013 

[

non applicable]

 

 

Electromagnetic disturbance allowance 

T

HE 

WIWE

 MOBILE AND HEART DIAGNOSTIC DEVICE WAS MADE TO BE USED IN AN ELECTROMAGNETIC ENVIRONMENT WITH THE CONDITIONS BELOW

.

 

 

T

HE USER NEEDS TO ENSURE

,

 THAT THE DEVICE IS FUNCTIONING WITH THESE CONDITIONS

.

 

 

Disturbance allowance examination 

IEC 60601 

examination level 

Compliance 

Electromagnetic environment - guideline 

Electrostatic discharge (ESD)  
MSZ EN 60601-1-2:2008 

± 6 kV 

on contact 

± 8 kV 

on strikeover 

approved 

Wood, concrete or ceramic floor case needed. If the floor is covered with 
plastic, then the relative humidity should be at least 30 %. 

Quick electrical transients/burst 
MSZ EN 60601-1-2:2008 

±2kV L, N–

reference 

between earth 

approved 

The energetic quality of the network needs to be equal with the usual 
used in commercial or hospital environment.  

Shockwave (Surge) 
MSZ EN 60601-1-2:2008 

1kV L-N 

approved 

The energetic quality of the network needs to be equal with the usual 
used in commercial or hospital environment. 

Voltage breaking, short time voltage outages 
fluctuations on the input supply wires 
MSZ EN 60601-1-2:2008 

100%, 60%, 30% 

1, 5, 25 periods 

approved 

The energetic quality of the network needs to be equal with the usual 
used in commercial or hospital environment. WIWE was made to use in 
battery mode. At least 50 measurements can be done with it. 

Magnetic field with network frequency  
(50/60 Hz) 
MSZ EN 60601-1-2:2008 

3 A/m 

non applicable 

There is no part in the WIWE which is sensitive for the network’s 

electromagnetic field. 

 

Electromagnetic disturbance allowance 

T

HE 

WIWE

 MOBILE AND HEART DIAGNOSTIC DEVICE WAS MADE TO BE USED IN AN ELECTROMAGNETIC ENVIRONMENT WITH THE CONDITIONS BELOW

.

 

 

T

HE USER NEEDS TO ENSURE

,

 THAT THE DEVICE IS FUNCTIONING WITH THESE CONDITIONS

.

 

 

Disturbance 

allowance 

examiation 

IEC 60601 1-2 

examination level 

Compliance 

Electromagnetic environment - directive 

Conducted RF  
MSZ EN 60601-1-
2:2008 

3Veff 
0,15

80MHz 

modulation: 
2Hz, 80% AM 

[3Veff] V 
approved 

The portable and mobile RF transmission deivces can only be used within a distance derived from 

a formula based on the transmitter’s frequency.

 

Suggested distance for clearance 

d=[3,5/V1]√P

 

d=[3,5/E1]√P 80 MHz –

 800 MHz 

d=[7/E1]√P 800 MHz –

 2,5 GHz 

where P is the highest capacity of the transmission output in Watts (W) and d is the suggested 
distance for clearance in meters (m). 
The field strength determined by local measurement, coming from the deplyoed RF 
transmitters 

a.

 

should be smaller than the compliance level in case of each frequency interval 

b.

 

Disturbance can occur in the environment of devices with this symbol here: 

Broadcasted RF 
MSZ EN 60601-1-
2:2008 

3 V/m 
3V/m (0,08-
2,5GHz) 
modulation: 
2Hz, 80% AM 

[3] V/m 
approved 
 

N

OTES

1.

 

I

N CASE OF 

80

 

MH

Z AND 

800

 

MH

Z THE HIGHER FREQUENCY INTERVAL 

(

UPPER

)

 NEEDS TO BE USED

.

 

 

2.

 

T

HESE ARE ONLY GUIDELINES

.

 

I

T IS POSSIBLE

,

 THAT ELECTROMAGNETIC SPREADING CAN

T BE USED IN EVERY SITUATION

,

 IT IS DEPENDENT ON THE OBJECTS IN THE ENVIRONMENT

,

 THE SPATIAL LOCATION OF THE PEOPLE AND THEIR ABSORBING AND 

REFLECTIVE CAPABILITES TOO

 

Suggested clearance distances between the portable and mobile RF telecommunication devices and the WIWE mobile heart diagnostic device

 

T

HE 

WIWE

 MOBILE HEART DIAGNOSTIC DEVICE WAS MADE TO USE IN SUCH ELECTROMAGNETIC ENVIRONMENT

,

 WHERE THE 

RF

 DISTURBANCES ARE UNDER CONTROL

.

 

T

HE BUYER OR USER OF THE 

WIWE

 

MOBILE HEART DIAGNOSTIC DEVICE CAN HELP TO PREVENT THE ELECTROMAGNETIC DISTURBANCES BY DETERMINING THE SMALLEST CLEARANCE DISTANCE BETWEEN THE TELECOMMUNCATION DEVICES 

(

TRANSMITTERS

)

 AND THE 

WIWE,

 DEPENDENT ON THE TELECOMMUNACTION DEVICE

S HIGHEST OUTPUT CAPACITY

.

 

The transmitter

’s

 highest given output 

capacity 

W

 

Clearance dist

ance based on the transmitter’s frequency

 (m

)

 

150 kHz 

 80 MHz 

d=[3,5/V

1

]√P

 

80 MHz 

 800 MHz 

d=[3,5/E

1

]√P

 

800 MHz 

 2,5 GHz 

d=[7/E

1

]√P

 

0,01 

0,12 

0,12 

0,23 

0,1 

0,37 

0,37 

0,74 

1,17 

1,17 

2,33 

10 

3,69 

3,69 

7,38 

100 

11,67 

11,67 

23,33 

In case of the transmitter with the highest output capacity that are not in this table the 

suggested clearance distance in meters (m) can be determined with the use of the equation dependent on the transmitter’s freq

uency, 

where P is the highest ouput transmitting capacity given buy the manufacturer in Watts (W). 
1.

 

N

OTE

: In case of 80 MHz and 800 MHz the higher frequency interval (upper) needs to be used 

2. N

OTE

:

 

These are only guidelines. It is possible, that el

ectromagnetic spreading can’t be used in every situation, it is dependent on the objects in the environment, the spatial loca

tion of the people and their absorbing and 

reflective capabilites too. 

 

Summary of Contents for DSW0001

Page 1: ...ver Not all cardiovascular diseases can be detected with WIWE so if you experience any symptoms of acute heart disease or if you are unsure of your symptoms regardless of the test results of WIWE plea...

Page 2: ...data The ECG curve appears in correct proportion on the smartphone s screen that is why it can be monitored without using any paper WIWE works in full funcionality with its application running on a sm...

Page 3: ...e application s Information section 5 3 Measurement After initiating a measurement on your smartphone press the ON button as the appearing illustration shows you to on your phone in order to establish...

Page 4: ...one is not available to the user In such cases there is also an opportunity to take a measurement In such cases please follow these steps 1 Turn on the device 2 Make sure that the failure warning sign...

Page 5: ...ectromagnetic compatibility Electronic appliances such as mobile telecommunication device radiation may disturb the operation of WIWE therefore always consider the signals indicating electromagnetic c...

Page 6: ...n phones with iOS 8 1 which have Bluetooth 4 0 and Bluetooth Low Energy protocol and standard List of compatible devices iPhone 4S iPhone 5 iPhone 5C iPhone 5S iPhone 6 iPhone 6 Plus Android The appli...

Page 7: ...nication Connection to a smartphone Bluetooth 4 0 Controls ON button 2 electrodes everything else from WIWE software via phone Accessories Standard micro USB cable Heart rate measurement range 30 240...

Page 8: ...the network s electromagnetic field Electromagnetic disturbance allowance THE WIWE MOBILE AND HEART DIAGNOSTIC DEVICE WAS MADE TO BE USED IN AN ELECTROMAGNETIC ENVIRONMENT WITH THE CONDITIONS BELOW T...

Reviews: