Things to do before using WIWE 

After unpacking WIWE, before usage, the mobile application required for operation needs to be downloaded and started. Once this is 
done and the device is turned on by pressing the button on the top of the device, WIWE – within the normal operational temperature 
threshold and on an adequate battery level – can be utilized right away. In case of extreme storage temperature you need to wait while 
the temperature of the device’s case reaches the normal operational threshold, which means approximately 10 minutes from -25 °C to 
+5 °C and approximately 5 minutes from +60 °C to +40 °C in case of an environment with normal room temperature. 

4.1

 

Ready for operation  

For the proper operation of WIWE, please check the conditions below before every usage: 

 

Intactness of the device. Make sure that the device is not damaged and in a usable condition! 

 

Cleanness of the sensors. 

 

Battery level. The device is fully charged when the 4 LED lights on top are lit when you turn on the device. If at least two of the 
LED lights are not lit, the accumulator must be charged before use. See Charging the accumulator section. 

5

 

Measuring 

Users can take a measurement with the help of the WIWE devie and the mobile application, during which the resulting measurements 
can be viewed on the phone screen and these can be shared with others (eg. with your doctor). 

5.1

 

Before measuring: 

Please consider the points below in order to get precise measurement results: 

 

Electrodes should be in contact with the skin of the fingertips directly. 

 

Fingertips should completely cover the electrodes. 

 

If your skin is very dry, make it moist with a wet cloth before measuring. 

 

If the electrodes are contaminated, remove the contamination from the surface with a soft wipe. 

 

When measuring, the hands should not touch any other body part. Please bear in mind that the skin of your left and right hand cannot 
touch during measuring, and neither the user’s skin, nor the device’s electrodes can touch metal devices. Otherwise measurement 
cannot be performed properly. 

 

Do not move during the measuring, because any muscle movement can cause false measurement. 

 

Measuring should be performed preferably in sitting or lying position, but not standing. 

5.2

 

Account registration 

 

Since WIWE and the application can be used by more than one user, medical data and measurements of the individuals are suggested 
to be separated. For this, you have the opportunity to create accounts on your mobile phone. 

 

Account registration cannot be skipped, however, there is a possibility to initiate measuring anonymously, without an account. 

 

You  can  find  detailed  information  about  registering  an  account  in  the  Advanced  guide  of  your  mobile  application’s  Information 
section. 

5.3

 

Measurement: 

 

After initiating a measurement on your smartphone, press the ON button as the appearing illustration shows you to on your phone in 
order to establish the radio (Bluetooth) connection.    

 

Measurement is done by placing the same fingers of your left and right hand on the sensors. 

 

Touch  the  electrodes  and  gently  hold  your  fingers  on  them.  For  a  successful 
measuring, a preparatory process is also required, when the user places their same 
two  fingers  of  each  of  their  hands  on  the  WIWE  sensors  and  becomes  relaxed. 
During  the  preparatory  period,  the  application  checks  whether  the  contact  was 
adequate,  and  whether  the  data  received  from  WIWE  are  suitable  for  starting  a 
measurement. The preparatory period is 10 seconds long.  

 

In case of a normal measurement the mobile phone, in individual cases the process 
indicator on the front of the WIWE shows that the measuring process has started. 

 

Letting  go  of  the  sensors  within  30  second  will  result  in  the  interruption  of 
measurement. 

 

After 60 seconds, the application indicates on the phone screen when data collection is finished. During data collection, the application 
displays the ECG, their pulse and blood oxygen saturation level of the user who touched the sensors. 

 

At the end of the measurement, evaluation may take a few seconds, then you can check the results on your mobile phone. 

 

In case of an unsuccessful measurement please follow the instructions appearing on your phone. Repeat the measurement. 

 

When  the  failure  signaling  LED  blinks  on  the  device,  it  means  that  ECG  signs  are  unstable  or  weak.  In  this  case  repeat  the 
measurement. 

 

The device will only start measuring, if the battery voltage level allows at least one measurement. 

5.4

 

Frequent causes of inaccurate measurement 
1.

 

Inadequate contact between the electrodes and the fingers. 

2.

 

Left hand skin and right hand skin touch during measuring. 

5.5

 

Results display screen 

After the measurement and evaluation, measurement data can be viewed on your phone screen. It is recommended to show these results 
to a doctor as well, so that any changes in your health may be diagnosed by them. 
To have a quick overview of all these data, measurement and evaluation data are summarized on the Evaluation screen. 

The evaluation of the main measured data is indicated by 3 circles in a row of the same 
diameter on the screen. In the center of the circles, ECG (Electrocardiography), AF (Atrial 
fibrillation)  and  VH (Ventricular  heterogeneity)  abbreviations  show  which  health 
parameter group they stand for. The filling colors of the circles can be green, yellow or 
red,  depending  on  the  evaluation  of  the  measurement’s  results.  Green  color  stands  for 
„normal” condition, yellow means „minor deviation from normal condition” and red is 
„major deviation from normal condition”. 
Underneath the circles, SPO

2

 button shows oxygen saturation, whereas Bpm shows the 

pulse value. 

 

Summary of Contents for DSW0001

Page 1: ...ver Not all cardiovascular diseases can be detected with WIWE so if you experience any symptoms of acute heart disease or if you are unsure of your symptoms regardless of the test results of WIWE plea...

Page 2: ...data The ECG curve appears in correct proportion on the smartphone s screen that is why it can be monitored without using any paper WIWE works in full funcionality with its application running on a sm...

Page 3: ...e application s Information section 5 3 Measurement After initiating a measurement on your smartphone press the ON button as the appearing illustration shows you to on your phone in order to establish...

Page 4: ...one is not available to the user In such cases there is also an opportunity to take a measurement In such cases please follow these steps 1 Turn on the device 2 Make sure that the failure warning sign...

Page 5: ...ectromagnetic compatibility Electronic appliances such as mobile telecommunication device radiation may disturb the operation of WIWE therefore always consider the signals indicating electromagnetic c...

Page 6: ...n phones with iOS 8 1 which have Bluetooth 4 0 and Bluetooth Low Energy protocol and standard List of compatible devices iPhone 4S iPhone 5 iPhone 5C iPhone 5S iPhone 6 iPhone 6 Plus Android The appli...

Page 7: ...nication Connection to a smartphone Bluetooth 4 0 Controls ON button 2 electrodes everything else from WIWE software via phone Accessories Standard micro USB cable Heart rate measurement range 30 240...

Page 8: ...the network s electromagnetic field Electromagnetic disturbance allowance THE WIWE MOBILE AND HEART DIAGNOSTIC DEVICE WAS MADE TO BE USED IN AN ELECTROMAGNETIC ENVIRONMENT WITH THE CONDITIONS BELOW T...

Reviews: