4.5 Line Voltage Switch
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.5 Line Voltage Switch
Tools required:
Phillips head screwdriver
Removal
SHOCK HAZARD:
Locate the involved building standard or emergency circuit breaker
panel. Set the involved circuit breaker to OFF. Lock out and tag out the
breaker. Failure to follow this procedure could cause serious personal
injury and damage to the equipment.
1. Set the involved circuit breaker (A) to OFF (see figure 4-1 on page 4-3).
2. Lock out and tag out the circuit breaker.
3. Using the phillips head screwdriver, remove the faceplate screws (C) and
the faceplate (B) (see figure 4-6 on page 4-13).
Figure 4-6. Line Voltage Switch Removal
m142_017
4
Integris® 2002 Headwall System Service Manual (man142rb)
Page 4 - 13
Summary of Contents for Integris 2002