4.1 Duplex Receptacle
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.1 Duplex Receptacle
Tools required:
Phillips head screwdriver
Removal
SHOCK HAZARD:
Locate the involved building standard or emergency circuit breaker
panel. Set the involved circuit breaker to OFF. Lock out and tag out the
breaker. Failure to follow this procedure could cause serious personal
injury and damage to the equipment.
1. Set the involved circuit breaker (A) to OFF (see figure 4-1 on page 4-3).
2. Lock out and tag out the circuit breaker.
Figure 4-1. Standard or Emergency Circuit Breaker Box
4
A
m142_012
3. Remove the screws (B) from the duplex receptacle faceplate (C) (see figure
Integris® 2002 Headwall System Service Manual (man142rb)
Page 4 - 3