Zhengzhou Winsen Electronics Technology Co., Ltd
www.winsen-sensor.com
Tel: 0086-371-67169097/67169670
Fax: 0086-371-60932988
Email: [email protected]
7. Anti-Interference
ME3-CL2 sensor also responds to other gases besides target gas. Below are the response characteristics of
interferential gases.
Gas
Concentration
ME3-Cl
2
H2S
15ppm
<-3ppm
CO
200ppm
0ppm
NO
35ppm
0ppm
SO2
5ppm
0
ppm
HCN
10ppm
0ppm
H2
400ppm
0.1ppm
C2H4
400ppm
0.1ppm
HCL
5ppm
0ppm
CO2
5%
0
NH3
20ppm
0.1ppm
8. Application Notes
Sensor shall Avoid organic solvent, coatings, medicine, oil and high concentration gases.
All ME Sensors shall not be encapsulated completely by resin materials, and shall not immerse in
pure oxygen environment, otherwise, it will damage the function of sensor.
All ME sensors shall not be applied in corrosive gas environment, or the sensor will be damaged.
Please test the sensitivity of gas sensors in clean atmosphere.
Sensors Shall be avoided to face the gas, which flow directly from front side.
To avoid to bend and break of pins.