Overview

Winmore optical dissolved oxygen sensor measures dissolved oxygen using the principle of
oxygen dynamic luminescence quenching technique. When blue light excites sensor film, the
sensing film emits red light. The phase difference between the blue excitation and returned red
emission is measured, and the result is used to calculate DO concentration.

Features

Membraneless, no electrodes design
No oxygen consumption
Flow Independent
Automatic temperature compensation
No H2S Interference
Low Drift, fast response, and more accurate
Low Maintenance, low operating cost
Easy sensor cap replacement
RS-485; MODBUS protocol compatible

Summary of Contents for W504-A

Page 1: ...Oxygen Sensor User Manual Yantai Winmore Trade Co Ltd Address 701 of Zhongrun Building No 1889 Jintan East Road Laishan Zone Yantai City Shandong China Email info winmoreltd com Mobile Whatsapp 008615...

Page 2: ...al Specifications 6 Installation 7 Part List 7 Before use 7 Sensor Installation 7 Calibration 8 Calibration Solution Preparation 8 Calibration take 2 point calibration as an example 8 Appliances and r...

Page 3: ...ns specified in this manual any fault and loss caused shall not be covered by the warranty and the manufacturer shall not bear any relevant responsibility If you have any questions please contact our...

Page 4: ...ng film emits red light The phase difference between the blue excitation and returned red emission is measured and the result is used to calculate DO concentration Features Membraneless no electrodes...

Page 5: ...Yantai Winmore Trade Co Ltd www winmoreltd com 5 Introduction Structure chart Dimension 26 x151 mm x L W504 Structure chart...

Page 6: ...Temperature Accuracy 0 2 Interface Support RS 485 MODBUS protocols Construction 3 4 inch NPT Power Requirements DC 6 12V 5 current 50mA Temperature Sensor Type NTC Sensor OD 26mm Sensor Length 177 5...

Page 7: ...ector of sensor cable should be screwed tightly to avoid moisture incursion 2 Do not use the sensor cable to pull the sensor It is required to install sensor in a secure and stable mounting bracket 3...

Page 8: ...sensor can be directly put into the air calibration accuracy will be slightly deviated Calibration take 2 point calibration as an example Restore user calibration data by default K 1 B 0 see modbus d...

Page 9: ...sensor You only need to connect the probe to the calibration software through the computer interface and you can easily complete the calibration of the sensor according to the software prompts Applia...

Page 10: ...inner surface If water vapor or dust gets inside the fluorescent cap the cleaning steps are as follows 1 Remove the fluorescent cap 2 Rinse the inner surface of the fluorescent cap with tap water 3 F...

Page 11: ...rift and it needs to be soaked in water for 48 hours before continuing to work Attention Protect the inner surface of the fluorescent cap from sun exposure Do not touch the fluorescent film with your...

Page 12: ...ERROR POSSIBLE CAUSE SOLUTION No data displayed on controller if sensors are connected Connection issue Reconnect sensor to controller Cable failure Contact customer service Bad sensor cap or loose s...

Page 13: ...atural disasters social unrest war published or unpublished terrorism civil war or any governmentforced 2 Damage caused due to improper use negligence accident or caused by the improper application an...

Page 14: ...Yantai Winmore Trade Co Ltd www winmoreltd com 14...