15.6" Medical Display User Manual
A long period of use
Due to the characteristics of LCD panel, displaying static images for extended periods, or
using the unit repeatedly in a high temperature/high humidity environments may cause
image smearing, burn-in, areas of which brightness is permanently changed, lines, or a
decrease in overall brightness.
In particular, continued displaying of an image smaller than the monitor screen, such as in a
different aspect ratio, may shorten the life of the unit.
Avoid displaying a still image for an extended period, or using the unit repeatedly in a high
temperature/high humidity environment such an airtight room, or around the outlet of an air
conditioner.
Moisture condensation
If the unit is suddenly taken from a cold to a warm location, or if ambient temperature suddenly
rises, moisture may form on the outer surface of the unit and/or inside of the unit. This is known as
condensation. If condensation occurs, turn off the unit and wait until the condensation clears
before operating the unit. Operating the unit while condensation is present may damage the unit.
Cleaning before cleaning
Be sure to disconnect the AC power cord from the AC outlet.
Cleaning the monitor
A material that withstands disinfection is used for the front protection plate of the medical use LCD
monitor. The protection plate surface is specially treated to reduce reflection of light. When
solvents such as benzene or thinner, or acid, alkaline or abrasive detergent or chemical cleaning
cloth are used for the protection plate surface/monitor surface, the performance of the monitor
may be impaired or the finish of the surface may be damaged. Take care with respect to the
following:
Clean the protection plate surface/monitor surface with a 50 to 70 v/v% concentration of
isopropyl alcohol or a 76.9 to 81.4 v/v% concentration of ethanol using a swab method.
Wipe the protection plate surface gently (wipe using less than 1 N force).
Stubborn stains may be removed with a soft cloth such as a cleaning cloth lightly
dampened with mild detergent solution using a swab method and then clean using the
above chemical solution.
Never use solvents such as benzene or thinner, or acid, alkaline or abrasive detergent, or
chemical cleaning cloth for cleaning or disinfection, as they will damage the protection plate
surface/monitor surface.
Do not use unnecessary force to rub the protection plate surface/monitor surface with a
stained cloth. The protection plate surface/monitor surface may be scratched.
Do not keep the protection plate surface/monitor surface in contact with a rubber or vinyl
resin product for a long period of time. The finish of the surface may deteriorate.