Chapter 4: Operating the Device
Guidance and
manufacturer’s declaration-electromagnetic emissions
The W15L100-PTA3-ME is intended for use in the electromagnetic environment specified below.
The customer or the user of the W15L100-PTA3-ME should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The W15L100-PTA3-ME uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The W15L100-PTA3-ME is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
D
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
Guidance and
manufacturer’s declaration - electromagnetic immunity
The W15L100-PTA3-ME is intended for use in the electromagnetic environment specified below. The customer or the
user of the M270TF-XXX / M320TF-XXX should assure that it is used in such as environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/ output lines
±2 kV for power supply lines
±1 kV for input/ output lines
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV differential mode
±2 kV common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
< 5%
(> 95% dip in) for 0.5 cycle
40%
(60% dip in) for 5 cycles
70%
(30% dip in) for 25 cycles
< 5%
(> 95% dip in) for 5 sec
< 5%
(> 95% dip in) for 0.5 cycle
40%
(60% dip in) for 5 cycles
70%
(30% dip in) for 25 cycles
< 5%
(> 95% dip in) for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the W15L100-PTA3-ME
requires continued operation
during power mains
interruptions, it is recommended
that the W15L100-PTA3-ME
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
NOTE: is the A.C. mains voltage prior to application of the test level.