WPC-766/WPC-766E User’s manual
8
Electrostatic
discharge
(ESD)
IEC
61000-4-2
8 kV
contact
15 kV air
8 kV
contact
15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least 30
%.
Electrical fast
transient/burst
IEC
61000-4-4
2 kV for
power
supply lines
1 kV for
input/output
lines
2 kV for
power
supply lines
1 kV for
input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC
61000-4-5
1 kV
line(s) to
line(s)
2 kV
line(s) to
earth
1 kV line(s)
to line(s)
2 kV line(s)
to earth
Mains power quality should be
that of a typical commercial or
hospital environment.