![background image](http://html1.mh-extra.com/html/wincomm/wmp-24c/wmp-24c_user-manual_984910023.webp)
WMP-24C/24F User’s manual
10
interruptions
and
voltage
variations
on power
supply
input lines
IEC
61000-4-11
0%
U
T
(100 % dip in
U
T)
for 0.5 cycle
0 %
U
T
(100 % dip in
U
T)
for 1 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
0 %
U
T
(100 % dip in
U
T)
for 250
cycles
0 %
U
T
(100 % dip in
U
T)
for 0.5 cycle
0 %
U
T
(100 % dip in
U
T)
for 1 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
0 %
U
T
(100 % dip in
U
T)
for 250 cycles
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
model
WMP-24C/24F
requires continued
operation during
power
mains interruptions, it
is recommended that
the model
WMP-24C/24F be
powered from an
uninterruptible power
supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE
U
T is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The model WMP-24C/24F is intended for use in the electromagnetic environment specified below. The
customer or the user of the model WMP-24C/24F should assure that it is used in such an environment.
Immunity
IEC
60601
test level
Compliance
level
Electromagnetic environment
–
guidance
Summary of Contents for WMP-24C
Page 27: ...WMP 24C 24F User s manual 14 Dimension WMP 24F VESA Mount Screw type M4...
Page 28: ...WMP 24C 24F User s manual 15...
Page 29: ...WMP 24C 24F User s manual 16...
Page 30: ...WMP 24C 24F User s manual 17 WMP 24C VESA Mount Screw type M4...
Page 31: ...WMP 24C 24F User s manual 18...
Page 32: ...WMP 24C 24F User s manual 19...
Page 63: ...WMP 24C 24F User s manual 50 J13 J15 DIMM1 DIMM0 DDR4 Vertical Connect...