Important: Please Read
For "Heat" and "Heat & Massage" Chairs: This equipment is marked with the industry recognized
symbol indicating it is a “Class II Double-Insulated” device as defined by IEC 60601-1. As such,
it does NOT require a three conductor power cord (i.e. one with a ground connection) for
safety purposes. While the equipment can be safely operated with no ground connection, a
three conductor power cord is provided to satisfy customer expectations for typical equipment
configuration. Devices qualifying as double insulated require different testing and examination
procedures than “Class I Grounded” devices. If your facility requires the product to be tested, be sure to follow
testing and examination procedures for “Class II Double-Insulated” devices.
ISOLATION DIAGRAM FOR HEAT & MASSAGE SYSTEM
Area
Insulation Type
A
Basic Operational
B
Double Insulated / Reinforced
C
Double Insulated / Reinforced
D
Double Insulated / Reinforced
E
Double Insulated / Reinforced
F
Double Insulated / Reinforced
G
Inlet FE Pin only connects to inlet & pad
Page 7 of 24