Step 3
Set up the 681XXB as follows for an ana-
log sweep power level flatness test:
a. Reset the instrument by pressing
SYSTEM
, then
Reset
. The CW
Menu is displayed.
b. Press
Analog
to place the 681XXB in
the analog sweep frequency mode and
display the Analog Sweep Menu.
c. With the Analog Sweep menu dis-
played, press the main menu key
FREQUENCY
CONTROL
The Sweep Frequency Control menu,
shown below, is displayed.
d. Press
Full
to select a full range fre-
quency sweep.
e. Press
Edit L1
to open the current
power level parameter for editing.
f. Set L1 to the power level noted on the
test record.
g. Now, return to the Analog Sweep menu
by pressing the main menu key
CW/SWEEP
SELECT
OPERATION
POWER LEVEL ACCURACY
VERIFICATION
AND FLATNESS TESTS
681XXB OM
5-15