Model P703B Operating Instructions
Use LA704B Line Adaptor to power unit from line voltage without batteries installed or
to charge NiCad batteries.
CAUTION: DO NOT ATTEMPT TO RECHARGE ALKALINES WITH THE LA704B.
Alkaline batteries may EXPLODE or leak corrosive fluids.
Test for Proper Operation
To check the Model P703B for proper
operation:
•
Use a digital multimeter to verify that the
proper voltage and current are available at
the transducer connector.
•
Substitute an Oscillator for the transducer.
•
Follow the connection to the power unit as
shown at right.
•
The unit should have unity gain.
—FOR TECHNICAL ASSISTANCE—
Please contact your Applications Engineer at Wilcoxon Research, Inc.
1-800-WILCOXON • 301-330-8811 • Fax 301-330-8873