20
WIKA operating instructions, models F2802, F2808
ADPR1X914114.01 07/2020 EN/DE
EN
9. Specifications
9. Specifications
Model F2802
Rated force F
nom
kN
0.5, 1, 2, 5, 10, 20, 30, 50
Relative linearity error d
lin
Steel
±0.03 % F
nom
Stainless steel
±0.05 % F
nom
Relative creep, 30 min.
Steel
±0.03 % F
nom
Stainless steel
±0.05 % F
nom
Relative reversibility error v
Steel
±0.03 % F
nom
Stainless steel
±0.05 % F
nom
Relative span in unchanged mounting
position b
rg
Steel
±0.03 % F
nom
Stainless steel
±0.05 % F
nom
Relative deviation of zero signal d
S, 0
±2 % F
nom
Temperature effect on the zero signal TK
0
≤ ±0.025 %/10 °C
Temperature effect on the characteristic
value TK
C
≤ ±0.025 %/10 °C
Limit force F
L
150 % F
nom
Breaking force F
B
200 % F
nom
Material of the measuring body
Stainless steel
Rated temperature range B
T, nom
-10 … +60 °C
Service temperature range B
T, G
-20 … +80 °C
Input resistance R
e
385 ± 30 Ω
Output resistance R
a
350 ± 5 Ω
Insulation resistance R
is
≥ 5,000 MΩ/DC 100 V
Output signal (rated characteristic value)
C
nom
2.0 ± 5 % mV/V
Electrical connection
Measuring cable ∅ 5 x 3,000 mm
Excitation voltage B
U, nom
DC 10 V (max. 15 V)
Ingress protection (per IEC/EN 60529)
IP65 (< 5 kN), IP67 (≥ 5 kN)
Weight in kg
0.5 kN
0.3
1, 2 kN
0.5
50, 10 kN
0.5
20, 30 kN
1.3
50 kN
1.4