WIKA DW03 Series Instruction Manual Download Page 3

 

INSTRUCTION MANUAL 

NI-210WE 

Rev. 4   04/20 

 

3 di 6 

The test instrument should have a measurement range 

approximately equal to or slightly wider than the pressure switch 

range and should have an accuracy consistent with the accuracy 

required to calibrate the set point. 

For example DW03 range 0,7/16mbar the accuracy of the test 

gauge must be ± 0,04 mbar to calibrate the set point with an 

accuracy of ± 0,16 mbar (1% of the adjustable span). 

Fig. 3 – Calibration circuit for instrument with adjustable range less 

than 60 mbar 

 

1- Pressure switch 

2- Test 

gauge 

3- Inlet valve 

4- Volumetric 

pump 

 

5- Capacity 

6- Outlet 

valve 

6.1 PRELIMINARY 

OPERATIONS 

Remove the blocking device fixed to the side of the instrument 

case (Fig. 4). Remove the cover by rotating it in an anticlockwise 

direction. 

Fig. 4 - Pressure switch blocking device 

 

a) Blocking 

gap 

b) Blocking 

bracket 

c) Blocking 

nut 

d) Plumbing 

wire 

6.2  CALIBRATION CIRCUIT AND OPERATIONS 

Prepare the calibration circuit as indicated in Fig.3. 

The warning lamps should be connected to contact in the NO or 

NC position according to the required contact action. 

Connection of C and NO terminals 

• If the circuit is open at the working pressure, the switch 

closes

 

the circuit as the pressure 

increases

 when the desired value is 

reached. 

• If the circuit is closed at the working pressure, the switch 

opens

 

the circuit as the pressure 

decreases

 when the desired value is 

reached. 

Connection of C and NC terminals 

• 

If the circuit is closed at the working pressure, the switch 

opens

 

the circuit as the pressure 

increases

 when the desired value is 

reached. 

• If the circuit is open at the working pressure, the switch 

closes

 

the circuit as the pressure 

decreases

 when the desired value is 

reached. 

The pressure switch must be mounted in the normal installation 

position, i.e. with the pressure connection pointing downwards. 

The pressure conn (or H) of the instrument must be 

connected to the pressure source and the connection - (or L) has 

to be left to the atmosphere 

Increase the pressure in the circuit up to the desired set point 

value for the first microswitch. Use a wide bladed screwdriver, as 

indicated on the label, turn the screw until the relative lamp turns 

on (or turns off). 

6.3  CHECK OF SET POINT 

Generate the normal working pressure and wait the pressure 

stabilisation. Vary the pressure into the circuit and record the set 

point value. Write the set point values on the adhesive label. 

Note

: the repeatability should be checked verifying for three 

times the set point (Pi) starting always from the same pressure 

value (Pw). The pressure cycle should be slowly to give the 

possibility to record the set point with enough accuracy. 

6.4 FINAL 

OPERATIONS 

Disconnect the instrument from the calibration circuit. 

Take the cover, ensure that the sealing gasket is correctly fitted 

into its seat, and insert the cover onto the case, with the blocking 

gap positioned in correspondence to the blocking bracket. 

Turn the cover clockwise closing it tightly. Mount the blocking 

device as in Fig. 4.  

Mount on pressure connection, cable entry and adjustment screw 

the protection caps supplied with the instrument.  

Caution:

 The protection caps should only be definitively 

removed 

during

 the connection steps. 

MOUNTING AND CONNECTIONS 

7.1 MOUNTING 

Surface

 mount the instrument by means of the holes provided, 

(see Fig.10 and 11). In case of surface or panel or rack mounting 

the instruments can be mounted side by side (see Fig.12 & 13). 

The chosen position must be such that vibrations, the possibility 

of shocks or temperature changes are within tolerable limits. The 

instrument 

must

 be positioned higher than the pipe inlet (Fig.14).  

7.2 PRESSURE 

CONNECTIONS 

Connecting lines are an integral part of the instrument in 

transmitting the measured variable from the measuring point to 

the instrument. 

The diameter and length of the two connecting lines between 

instrument and pressure taps are to be such as not to cause 

dampening or lag in transmitting the differential pressure. 

Run connecting lines always sloping down from instrument to 

process so that possible condensate flows towards process 

(avoid forming of siphons). 
For a correct installation (see Fig 14) it is necessary to: 

Mount

 a shut-off valve with drain (root valve) on each process 

pipe inlet to allow the instrument to be excluded and the 

connection tubing to be drained. It is recommended that said 

valve has a capstan blocking device aimed at preventing it being 

activated casually and without authorisation. 

Mount

 a 

3 valve manifold 

near the instrument to permit possible 

functional verification on site and removal of the instrument. It is 

recommended that the manifold is made up of two service valves, 

one by-pass valve and two suitably connected drain plugs. The 

three valves with the drains can be reunited by a single device 

called a “Three valve manifold”  

Mount

 a three piece joint onto the threaded attachment of the 

instrument to permit the easy mounting or removal of the 

instrument itself. 

Carry out

 the connection using a flexible tube in such a way that 

variations in the temperature of the tube itself do not force the 

instrument attachment. 

Ensure

 that all the pressure connections are airtight. It is 

important that there are no leakage in the circuit. 

Close root valves, the two service valves, drain plugs and open 

the by-pass valve. 

 

Reviews: