WIKA operating instructions digital indicator, model DI35-D
EN
7.8 Setting the switching outputs
The switching outputs can be configured independently. Each switching output can be
assigned a separate measuring input. The switching outputs react to the measuring
input at any time, regardless of which measuring input is shown on the digital display.
7.8.1 Flashing display on reaching the switch points
A switch point being exceeded can be signalled via a simultaneous flashing of the
7-segment display and switch point display.
PN
Function
Parameter
59
Flashing display
0 = not flashing
1 = at switch point 1
2 = at switch point 2
3 = at switch point 3
4 = at switch point 4
5 = at switch point 1 + 2
6 = at switch point 3 + 4
7 = at switch point 1 + 2 + 3 + 4
Reading the switching status
The switching status of the switching outputs can be read from the switch point display.
7.8.2 Reference value
The reference value is the value to which the switching output reacts.
PN
Function
Parameter
60
Reference value, switching output 1
0 = Deactivated
1 = Measuring input 1
2 = Measuring input 2
3 = Arithmetic function
70
Reference value, switching output 2
80
Reference value, switching output 3
90
Reference value, switching output 4
Switch point display