6. Commissioning, operation
6.4.6.2 Configure existing routines
While pressing on a created routine on the left side under the menu item
[Configure existing routine]
all settings are shown
on the right side. There changes can be done as described in chapter 6.4.6.1 “New calibration routine” and stored for the
future.
6.4.6.3 Start the routine
To start the calibration routine only press the button
[▶]
on the right side of the selected calibration routine.
Before starting the programme ensure the following points:
1. The calibrator is connected to the CTR3000 via the front-end USB before switching on the CTR3000.
⇒
Please use the delivered interface cables from WIKA. Otherwise we cannot guarantee the functionality.
2. The calibrator is switched on and is running.
3. The CTR3000 is switched now.
Please ensure that the right probes are assigned to the channels due, the CTR3000 is reading the probe
information of each channel during the calibration.