20
WIKA operating instructions, model CPH6200
EN
11221780.03 07/2022 EN/DE
6.3 Menu functions
Menu Para-
meter
Values
Meaning
[MENU]
▶
▲
or
▼
SEt
ConF
Set configuration: General settings
Unit
mbar, bar, ...
Display of unit
1)
2)
SL
oFF
/
on
Sea level: Switch altitude correction on or off
1)
2)
ALti
-200 ... +9999
Altitude: Altitude correction [metre] (only if SL is activated)
1)
2)
rAtE
SLo
Slow: Slow measurement (4 Hz filtered, low current
consumption)
1)
FASt
Fast: Fast measurement, filtered (> 1,000 Hz)
1)
P
.
dEt
Peak detection: Fast measuring rate, unfiltered (> 1,000 Hz)
1)
t
.
AU6
1 ... 120
Time in seconds, calculated via the averaging function
3)
oFF
Averaging deactivated
3)
P
.
oFF
1 ... 120
Auto power-off delay in minutes.
If no button is pressed and there is no data transfer via the
interface, the instrument will switch itself off automatically after
this interval.
oFF
No output function, lowest current consumption
Ovt
oFF
Instrument output is serial interface
SEr
Instrument output is analogue output 0 ... 1 V
dAC
Instrument output is serial interface
Adr
.
01, 11 ... 91
Base address of the interface (only with
Out
=
SEr
)
dAC
.
CH1, CH2 or
CH DIF
Measuring input which should be used for the analogue output
(only with
Out
=
dAC
)
dAC
.
0
Sensor dep.,
e.g. -5.00 ...
+5.00 mbar
Zero point setting for
Out
=
dAC
: Input of the measured
value at which the analogue output should output 0 V
(only for
Out
=
dAC
)
dAC
.
1
Sensor dep.,
e.g. -5.00 ...
+5.00 mbar
Slope setting for
Out
=
dAC
: Input of the measured
value at which the analogue output should output 1 V
(only for
Out
=
dAC
)
5. Commissioning, operation